F-751 Avocado Operation Manual Rev. 1/25/2023
Page
|
25
Felix Instruments
1554 NE 3rd Ave
Camas, WA 98607, USA
Phone: +1 (360) 833-8835
Fax: +1 (360) 833-1914
www.felixinstruments.com
[email protected]
Updating App
Note: Offset calibration is always recommended after installing new
firmware or before using new apps. The offset calibration procedure is
. Consult Felix Instruments support with any
questions.
1.
Determine the app to be installed is compatible with the firmware
version installed on your device. Refer to the Product Support >
Downloads page compatibility tables.
2.
Before performing any installation process, it is recommended to
backup all files that reside on the SD Card onto your computer.
a.
Power ON the F-75X and ensure all the measurements on
the device are archived to the SD Card.
b.
Power OFF the device, remove the SD Card and copy the
entire contents of the SD Card onto your computer.
3.
With the SD Card still attached to your computer, delete the old
app(s) from the SD Card. This step is optional, it is allowable to
have multiple apps on the SD Card however the following
considerations apply:
a.
If more than one app resides on the SD Card, the F-751
will power on to the “Ready” screen. The user will then
be required to press the right arrow button to load the
desired app.
b.
If only one app resides on the SD Card, the unit will load
the single app immediately when powered on and no
further action will be required. It is recommended that
the F-751 only have one app on the SD Card.
4.
Copy the new app onto the SD Card.
5.
Insert the SD Card back into the F-751.