Step-by-Step Assembly:

[email protected]

Page 9

Adjusting the Field Iris Diaphragm:

See image on the bottom of page 8 next to 

“Centering the Condenser” for reference.

By limiting the diameter of the beam of light 

entering the condenser, the field iris diaphragm (2) 

can improve the contrast of the image. 

Aperture Diaphragm Adjustment:

The aperture diaphragm dictates the numerical 

aperture (NA) of the illumination.  

The best resolution, contrast, and depth of field is 

obtained when the NA of the illumination matches 

with the NA of the objective.

Rotate the diaphragm adjusment ring (3) so the 

pointer matches up with the inscription (4) of the 

objective being used.

Using the 100x Immersion Oil Objective:

Start with the 4x objective and focus on the sample.

Place a drop of immersion oil (1) on the specimen.

Rotate the nosepiece counterclockwise so the 100x 

objective is in the light path and the oil seals the 

space between the objective and the cover slip.

Use the fine focus knob to obtain a clear image.

Make sure there are no air bubbles in the oil. If there 

are, rotate the nosepiece slightly to remove them.

Open the aperture diaphragm and field iris 

diaphragm fully.

After use, wipe the lens with a tissue moistened with 

microscope cleaning solution.

Using Filters:

Filters can increase the contrast in microscopy 

images.

The filter sits on top of the light as shown at right.

The RB30 accepts a 45mm diamter filter.

Summary of Contents for RB30

Page 1: ...RB30 Microscope User s Manual info feinoptic com...

Page 2: ...info feinoptic com Page 1 Microscope Components...

Page 3: ...igh temperatures dust or damp environments Use a flat work surface Set the power switch to off O before replacing a bulb or fuse and wait until the lamp is cool The power switch 4 is located on the si...

Page 4: ...he detergent is removed from the frame with a clean damp cloth prior to drying the surface Do not disassemble the microscope After use cover the microscope with a dust cover and power off the light Ob...

Page 5: ...Binocular Trinocular Microscope Assembly info feinoptic com Page 4...

Page 6: ...Fluorescence Microscope Assembly info feinoptic com Page 5...

Page 7: ...operating the microscope always start using the lowest objective 4x or 10x to get the sample into focus before moving to a higher magnification Eyepiece Assembly Remove the cover from the eyetube 1 In...

Page 8: ...e brightness and counterclockwise to decrease brightness Using the bulb at a lower brightness extends bulb life Slide Placement Push the specimen holder 1 back Place the slide 2 in the slot with the c...

Page 9: ...until the image is clear If multiple people are using the microscope note your diopter setting for quick transition from users Interpupillary Distance Adjustment When observing with both eyes adjust t...

Page 10: ...p with the inscription 4 of the objective being used Using the 100x Immersion Oil Objective Start with the 4x objective and focus on the sample Place a drop of immersion oil 1 on the specimen Rotate t...

Page 11: ...eyepieces are in focus adjust the focus on the c mount with an allen key to parfocal the camera with the eyepieces Phase Contrast Slider Assembly Insert the slider 1 from left to right with the inscri...

Page 12: ...move one eyepiece and insert the phase centering telescope Set the phase adjustment ring 2 to correspond with the phase objective being used example PH10x with 10x objective Adjust the centering teles...

Page 13: ...the binocular trinocular head on top of the fluorescence module and tighten the set screw 3 Insert transformer 1 into socket 2 and then plug directly into outlet Fluorescence Light Adjustment Fluores...

Page 14: ...r glass thickness 0 17mm Cover slip is on the bottom of the slide Put cover glass face up Immersion oil has dried on objective lens Clean objective lens No oil immersion was used with 100x lens Use im...

Page 15: ...ched properly Adjust specimen holder Stage is not fastened properly Reattach the stage Power is not turned on Check power cable connection LED light is not inserted properly Check light connection LED...

Page 16: ...to each 3 Disconnect the wires from the bulb assembly and unscrew the bulb assembly from the bulb compartment 4 Solder the wires to the new bulb assembly in the same location as the old bulb assembly...

Reviews: