IRB20 Inverted Microscope Bulb & Fuse Replacement
8
Fein
When you lift the bulb housing be careful
since the wires are still connected. In order
to replace the bulb remove the solder joints
from the wires connected to the bulb.
Unscrew the 2 screws holding the bulb to
the housing. Screw the new bulb into the
housing. Re-solder the wires back onto the
bulb in the same positions they were in
prior to bulb replacement.
Replacement Bulb: M40-001 (5W LED).
FUSE REPLACEMENT:
Turn off and unplug the microscope prior to fuse replacement.
The fuse box is installed below the power socket.
Fuse: 5x20mm, 2A, 250V.