background image

FZ12 Microscope Assembly

5

Fein 

3

5

6

7

1

2

8

9

4

Summary of Contents for FZ12

Page 1: ...Fein FZ12 Microscope User s Manual info feinoptic com...

Page 2: ...Fein FZ12 Microscope Components Beam Splitter Fine Focusing Knob Main Objective Aperture Diaphragm Adjusting Knob Coarse Focusing Knob BF DF Reflector Knob Zoom Knob Tension Adjustment Knob FZ12 BFDF...

Page 3: ...2 Fein FZ12 Microscope Components Lock Knob Focus Holder Microscope Body Trinocular Head Eyepieces Stage Clips Stage Plate FZ12 BFDF...

Page 4: ...2 Microscope Components Trinocular Head Eyepieces Focus Holder Fine Focusing Knob Coarse Focusing Knob Aperture Diaphragm Adjusting Knob LCD Display Intensity and Color Temperature Color Temperature C...

Page 5: ...4 Fein FZ12 Microscope Components Zoom Knob Click Stop Rheostat Control Power On Off Main Objective FZ12 ILB Ultra Thin base Microscope Head...

Page 6: ...FZ12 Microscope Assembly 5 Fein 3 5 6 7 1 2 8 9 4...

Page 7: ...focus bracket group with a M4 inner hexagon Allen key Replace the decorative cover 1 Assemble the Objective Attach the objective 2 into the bottom of the microscope body 1 by rotating the objective un...

Page 8: ...he eyepieces touches the bottom of the eyepiece tube Secure the eyepieces with the set screws 2 Assemble the Stage Plate Place the stage plate into the well space on the microscope stand To remove the...

Page 9: ...rce Make sure the main switch is at the O off position Insert one end of the optical fiber 1 into the inner hole 2 of the power supply box aligning the flat surface with the lock screw 3 Secure the op...

Page 10: ...d power off the light For a clear image ensure you do not leave fingerprints on the eyepieces or auxiliary lenses Wipe lenses gently with a soft tissue Wipe off fingerprints from lens surfaces with le...

Page 11: ...ward and backward to achieve different lighting effects 1 2 1 Adjust the Focusing Tension If the handle is very heavy when coarse focusing or the specimen leaves the focus plane soon after focusing or...

Page 12: ...djustment until it is clear Look through the left eyepiece If the image is not clear adjust the left diopter adjustment to get a clear image Left diopter adjusting ring Right diopter adjusting ring Ad...

Page 13: ...se to disengage the detent function Do not over rotate the positioning screw as it will damage the hosing and internal mechanisms of the microscope Click Click Stop 1 Using the Eye cups The eyepieces...

Page 14: ...o the trinocular tube and secure in place with the lock screw 1 Connect the camera to the c mount adapter Ensure the beam splitter pole 4 is in the outermost position for observation using the trinocu...

Page 15: ...is poor and details are poorly visible Objective lens is dirty Clean objective lens Dirt or dust on eyepieces Clean eyepieces Field of view in one eye does not match that of the other Interpupillary...

Reviews: