Recommended separation distances between
portable and mobile RF communications equipment and the Wrist Blood Pressure Monitor.
The
Wrist Blood Pressure Monitor
i
s
intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the
Wrist Blood Pressure Monitor
can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the
Wrist Blood Pressure Monitor
as recommended below, according to the maximum output power of the communications
equipment.
Rated maximum
output power of
transmitter (W)
Separation distance according to frequency of transmitter (m)
150 KHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.5 GHz
0.01
0.117
0.117
0.233
0.1
0.369
0.369
0.738
1
1.167
1.167
2.333