Calibration Parameters Verification
The instrument uses a set of predetermined calibration parameters to process flow signals. Most of these parameters should not
change. A data package shipped with the instrument (typically on a CD-ROM) contains the
ST75/ST75V Delta R Data Sheet
. This
contains the calibration parameters stored in the flow transmitter at the factory. To verify that these parameters have not changed,
complete the following:
1. Identify the appropriate Delta R data sheets by serial number of the instrument.
2. Press [D] [ENTER] to examine each of the parameters. The [ENTER] key allows scrolling one message at a time. Use Table
4 below to record actual instrument parameters. Compare with the Delta R data sheet’s ST75/ST75V parameters.
Table 4. Diagnostic Test Sequence on Display
An issue may exist if parameters with an asterisk (*) have changed. Contact Customer Service if this is the case. If the parameters
have not changed, continue with the next section.
S/W Version:
dR Min:
T SpanIDA C 0:
Flow Factor:
dR Max:
T ZeroIDA C 0:
Cmin Flow:
Cal Ref:
T SpanIDA C 1:
Cmax Flow:
Tcslp:
T ZeroIDA C 1:
Eng Units :
Tcslp 0:
State 0:
Line Size 0:
Tcslp 2:
Switch Pt 0:
Line Size 1:
Tot Menu:
State 1:
Cmin Temp :
Tot Flag:
Switch Pt 1:
Cmax Temp :
Totalizer:
K factor 1:
Min Flow:
Rollover Cnt:
K factor 2:
Max Flow:
Fix Pt Flag:
K factor 3:
Density:
Pulse Factor:
K factor 4:
*C1 [1]:
Pulse Out:
I factor:
*C1 [2]:
Hours:
Temp Flag:
*C1 [3]:
Sample Period:
Out Mode:
*C1 [4]:
dR Slope :
Namurmode:
*C1 [5]:
dR Off Set :
Boxcar Max:
Break Pt:
Refr Slope:
RTD-SLP-385:
*C2 [1]:
Refr Off Set:
% of Range:
*C2 [2]:
SpanIDAC 0:
User Name:
*C2 [3]:
ZeroIDAC 0:
Shop Order #:
*C2 [4]:
SpanIDAC 1:
Serial No.:
*C2 [5]:
ZeroIDAC 1:
Mode l#: