27
1. Device Settings
This will take the user to submenu with 4 options.
a.
Set Locations:
b.
Set Ch0 SN
c.
Set Ch1 SN
d.
Set Ch2 SN
1.a. Set Locations is not yet implemented, but will allow the user to set any
Ch ID to a location. Currently only the RS232 Serial unit can be Ch0.
1.b. Set Ch0 SN: Use this to set the Serial Number of the T-HRECS device on Ch0.
1.c. Ch0 SN: Use this to set the Serial Number of the T-HRECS device on Ch1.
1.d. Set Ch2 SN: Use this to set the Serial Number of the T-HRECS device on Ch2.
2. Station Settings
Select From txt list to read the “station_list.txt” file from the SD Card and pre-
pend a station name to each file before each SDIQ.
Select From csv db to read the “station_list.csv” file from the SD Card and pre-
pend a station name to each file before each SDIQ, and also read the station
and project ID, and Ch0, Ch1, and Ch2 CFT values from the SD Card.
3. Set Time
Rotate the dial to set the time.
4. Set Date
Rotate the dial to set the date.
5. SDIQ Settings
Select to access secondary menu related to the SDIQ.
The SDIQ Settings Menu consists of 13 options:
a.
Mass (kg)
b.
CFT0(mg/l)/(uS/cm)
c.
CFT1 (mg/l)/(uS/cm)
d.
CFT2 (mg/l)/(uS/cm)
e.
Inj. Conc. [mg/l]
f.
Inj. Volume (l)
g.
Stream Sample (l)
h.
BG EC.T N
i.
Set Interval
j.
Grade A Thresh %
k.
Grade B Thresh %
l.
Log Interval (sec)
m. Restore Defaults
n.
Back
5.a. Mass (kg): The default mass for each SDIQ. Factory Default is 2kg.
5.b. CFT0(mg/l)/(uS/cm): The default CF.T. Factory Default is 0.486(mg/l)/(uS/cm).
5.c. CFT1(mg/l)/(uS/cm): The default CF.T. Factory Default is 0.486(mg/l)/(uS/cm).
5.d. CFT1(mg/l)/(uS/cm): The default CF.T. Factory Default is 0.486(mg/l)/(uS/cm).
5.e. Solute EC.T (uS/cm): This value is used to correct the CF.T if a standard solu-
tion is used for CF.T derivation. Factory default is 5
µ
S/cm.
C H A P T E R 4 : Q IQUAC M E N U S YS T E M