DAZZLER
TM
system manual Part I : installation & operation
5.1
Figure 5.8: Warning messages prevent burnout when loading
When changing the operation mode from single to continuous or switching playable waveform
between A and B, an error message appears if excessive RF power would be applied (
The mode change will be aborted.
Figure 5.9: Warning messages prevent burnout when changing mode
Each wave sent to the generator contains meta data that gives information on the power damage
threshold and the minimum time period in which one cannot send another RF waveform to
the crystal without reaching this threshold. Any trigger occurring during this time period is
ignored and a missed trigger warning appears (
It is possible to visualize the signal corresponding to this minimum time period using the S2
or S4 generator output (
protectp
see
). This security prevents to send excessive
V3.00 - 8
th
April 2019