![FASTLITE DAZZLER Manual Download Page 37](http://html1.mh-extra.com/html/fastlite/dazzler/dazzler_manual_3588631037.webp)
DAZZLER
TM
system manual Part I : installation & operation
4.3
Figure 4.12: Pulse used to adjust the
Trigger to Laser
value.
4.3
Modes of operation
The word “mode” has been overloaded, it is used to loosely qualify various states and operation
modes:
1. CONTINOUS versus “Normal” or “single”
2. ONLINE versus OFFLINE
3. Alternate versus MemA or MemB “display”
4. streaming, table driven sequence mode
5. wave cycling
6.
Trig on Previous
versus “normal” triggering
7. No Delay
8. High Repetition Rate
4.3.1
Normal mode
Normal repetition mode is automatically selected when condition
is verified:
Trigger to Laser
>
=
Trigger Delay
+
T
Xtal
+
T
p
(4.1)
Or, in simpler words, the laser system provides a trigger signal
sufficently ahead
of the optic
pulse to allow synchronization with adjustment of a short delay (few
µs
). Signals in this mode
are represented on figure
Trigger
Propagation (T
p
)
Laser
Trigger Delay
Trigger to Laser
Acoustic signal
at input face
Figure 4.13: Normal repetition mode
Trigger to Laser
value must be smaller than the
triggerperiod
= (1
/RepRate
). If not, a
dialog box is displayed and the warning ”Missed Trig (delay)” appears in the
Trig&Mode
panel
message bar (
, see
for more details concerning this warning).
V3.00 - 8
th
April 2019