12
Lampada dentale
AL
YA
SAFETY rEGULATIOnS
- The function of the ALYA dental light is to illuminate the oral cavity of the
patient during dental operations.
- The light must be used by qualified staff.
- The articulated arm and head joint ensure that the light beam is correctly posi-
tioned. Do not overload the arms and joints with impact on the limit switches.
- Installation of the system must only be performed by specialized staff.
- Check that the power voltage, shown on the data plate, corresponds to the mains
voltage.
- Do not block the light beam for patients at risk (i.e. with eye conditions). Use
correct protection and precautions.
- Information plate with LED indications fixed to the rear arm.
- The dental light must be installed on a specific control and supply device, such
as a dental unit,
or with an electrical system that complies with standard
IEC 364-1 and “national regulations for installation of electrical systems in
areas for medical use”.
- The device must be installed with an omnipolar shut-off switch that disconnects
it from the mains and is conformant to IEC/EN 61058. It is necessary to install
a green signal light to indicate when the lamp is switched on.
- Do not carry out any maintenance work when the power is on; disconnect the
power cable from the mains beforehand.
- Do not introduce items or tools into slits in the light head that could touch live points.
- Patients at risk (e.g. children with eye diseases) must not stare at the light beam.
Always use appropriate protection devices and precautions. Faro suggests using
the protective eyeware BLUE-BAN or BABY BLUE-BAN (for children).
- Not suitable for installation in environments containing flammable gas.
- Do not spray detergents-disinfectants directly on the head.
- Inadequate performance of the system does not in any way compromise the
safety of the patient.
- For cleaning plastic parts of the lamp ALYA,
Do not uSE
detergents-
disinfectants
containing:
aMMonIuM
HYDrIXIDE
–
SoDIuM
HYDroXIDE – MEtHYlEnE CHlorIDE – MEtHYl alCoHol
. Failure
to comply with this indication might cause:
•
risk of breakage of plastic
parts
•
risk of structural breakdown of joints with potential fall of the articulated
arm. In case of doubt, please contact CUSTOMER CARE FARO.
- Do not carry out maintenance or replacement work on parts other than those
shown in the manual. Any operation not indicated in the manual could compro-
mise the safety aspect of the device.
- Disinfect surfaces with hydroalcoholic disinfectants.
- Do not leave small parts unattended or within the reach of exposed persons
(children); they could be potential sources of danger.
- The dentist is obliged to used the disposal protection sleeves on the handles of
the lamp or ensure that versions without this possibility are sterilized.
- Divide materials according to their type (e.g. iron, rubber, plastic.). When scra-
ping and disposing of the materials, comply with the current legislation of your
country, and if necessary use recognized and authorized specialist companies.
- The packaging of the light is suitable for protecting it from the penetration of
external agents.
- The system in its original packaging can be transported or kept in storage for a
period of 15 weeks if the following environmental conditions are observed:
- ambient temperature from –20° C to + 70°C
- relative humidity from 10% to 90%
- atmospheric pressure from 500 to 1060 mBar
- The system must be used in the following environmental conditions:
- temperature from 10° to 40°C
- relative humidity from 30 to 75%
- atmospheric pressure from 700 to 1060 mBar