EN
Error Codes
16
Wile moisture meter is equipped with error messages to guide the user to get the
most reliable results, as well as to ensure that the meter is in good condition.
Error code
Cause
Solution
Critical Error
The meter must be submitted to an autho-
rized service center
7
Weight not detected
Repeat test / Do weight test
8
Taring result is too high
Do weight test and check the scale
9
Weight measurement uncertain /
meter shook during the measure-
ment
Keep the meter stable,
measurement uncertainty has increased
10
Moisture measurement result stabi-
lize slowly
Happens to very wet samples
11
Weight measurement too uncertain
/ meter shook during the measure-
ment
Keep the meter stable, measurement uncer-
tainty has increased
12
Temperature of the sample lower
than 2 C / Measurement uncertainty
is high
The tester is most accurate when Rice and
tester temperature is between 16C and 32C
13
Temperature of the sample above
50 C / Measurement uncertainty is
high
The tester is most accurate when Rice and
tester temperature is between 16C and 32C
14
Large temp. diff. between the Rice
and test cell / Measurement uncer-
tainty has increased
For better accuracy, Rice and meter test cell
should be at similar temperatures
15
Not on a suffciently level surface,
meter warns with sound alarm
Keep meter stable during measurement
16
Weight result is over 330 g
Check that leveling handle did not stay
above the test cell/ Do weight test
17
Weigh result is under 50 g
Check that leveling handle did not stay
above the test cell/ Do weight test
18
Hectoliter weight is 20% higher than
average
Measurement uncertainty has increased, Do
weight test
19
Hectoliter weight is 20% lower than
average
Measurement uncertainty has increased,
Do weight test
20
Test cell is not cleared before the
measurement
Clear test cell