This will display a waveform similar to the one shown below. It may take up to a
minute for the waveform to be displayed since the software waits for the next pulse
to occur. Note that the amplitude of the first waveform peak must be greater than
20 or an error will occur. If the waveform amplitude is less than 20, the instrument
may have a faulty cable or ultrasonic transducer or the ultrasonic transducers may
not have enough ultrasonic couplant (grease).
5.7
Stopping a Test
To stop the current test, the user must press the STOP TEST button on the RUN-
TIME MENU screen. Once the test has been stopped, no further data will be
logged and no further temperature control will be provided. You may view the last
test or any test stored on the UCA by pressing the VIEW TEST button and
selecting the appropriate file name.
5.8
Using the USB Memory Module
The UCA operates under a version of the Windows
®
operating system. This allows
the use of certain Universal Serial Bus (USB) peripherals, such as USB flash
memory modules. These USB memory modules may be used to archive or move
UCA test files to another computer. The USB memory modules may be used
without special drivers on any Windows
®
XP or Windows
®
2000 based computers.
The memory module will simply appear as another disk drive when inserted into
the USB port on the computer.
To archive the stored tests on the UCA to the memory module, follow the steps
below.
1.
Connect the memory module to the USB port on the UCA control box. No