5.4
Instrument Setup Menu
From this menu the user can run diagnostics on the ultrasonic transducer circuitry
to determine if everything is functioning properly. This menu is also used to move
data files from the UCA computer to another PC in order to free space in the UCA
memory.
ARCHIVE DATA FILES – Select this command to move tests from the UCA to
the USB memory module. If the instrument reports a “Disk Full” message, it will
be necessary to do this before additional tests may be run. After this step is
completed, the stored tests on the UCA are moved then deleted. After this button is
pressed a message box will appear indicating which files were copied from the
UCA to the storage device.
CALIBRATE – The Calibrate button gives the operator the option of calibrating
temperature, pressure, or transit time.
To calibrate the transit time, the cylinder must be COMPLETELY full of water
with no air bubbles inside the cell.