background image

 

 

 

 

 

 

 

 

 
 
 

4. 

Grasp  the  flats  on  lock  collet  [64]  with  needle  nose  pliers  between  lower  spring 
coils.   Holding the dial [60] turn it, torsion shaft, and collet [65] counterclockwise to 
tighten the thin notched portion of lock collet [64] against spring [67]. and collet 67-
3]. 

 
C. 

Mounting Torsion Shaft Assembly in the Calibration Fixture. 

 
 

Re-assemble  the  Calibration  Fixture  and  set  it  up  for  the  Model  280  torsion  shaft 
assembly.

 

(Refer to Fig. 2). 

 
 

1. 

Install the pulley [3] into the top hole of the Pulley Support  Bracket.  

 
 

2. 

Loosen the two screws holding the lower bearing support bar, [1] and position it at 
the top of the slots for the 280 Rheometer and re-tighten the screws.  

 
 

3. 

Slide torsion shaft assembly into fixture from the top. Position torsion body hub into 
the hole in the top of the fixture and bearing [68] in the hole in the lower bearing 
support bar [1]. Rotate the torsion body to align the flat of the torsion body cross 
wise of the fixture.   Secure the torsion assembly into the fixture by locking the lock 
bar [4] over the flat of the torsion body.   

 

 

 

4.  

Remount the Dial [60] onto the top of the torsion shaft [66] with screw [79]. 

 
 

5. 

Position  the  hairline/magnifier  [2]  in  the  lower  groove  of  its  support.    Rotate  the 
hairline/magnifier [2] so the hairline aligns with the divisions on the dial and lock in 
place.  Loosen set screw [81-B] to loosen collet [67-2] in the torsion body [61], then 
rotate the dial and bob shaft to align Zero with the hairline.  Lock collet [67-2] in this 
position with set screw [81-B].  

 
 

6. 

Screw Calibration spool [5] onto bob shaft, threads down, until it is tight.  Wind two 
to three turns of thread around the drum clockwise as viewed from the top of the 
spool.  (Refer to Fig. 2.)  Route the thread over the sheave, and allow it to hang. 

 
D. 

Taking the Readings    

 
 

Hang the 25, 50 and 75 Gram weights each in turn on the loop in the end if the thread and 
note the dial readings obtained.  (Refer to Table 1).  If the dial readings are higher than 
shown in Table 1 - shorten the active part of the spring.  If the dial readings are lower than 
shown in Table 1  - lengthen the active part of the spring.   

 
E. 

Adjusting Torsion Spring: 

 
 

1. 

Loosen hex set screw [81-B] holding upper spring collet [67-2] in torsion assembly 
body [61]. 

 
 

2. 

Loosen torsion spring adjusting collet as described in Section 3B-2. If the readings 
were too high, screw the threaded outer collet [67-3] up the spring pairs (1/4 turn 
will  drop  3-5  dial  units).    If  the  readings  were  too  low,  screw  the  threaded  outer 
collet [67-3] down the spring pairs (1/4 turn will change reading 3-5 dial units). 

 
 

3. 

Align  top  of  outer  collet  [67]  slightly  above  top  of  widest  section  of  lock  bushing 
[65].  Tighten torsion spring adjusting collet as described in Section 3B-4.    

 

Summary of Contents for 280

Page 1: ...Rheometer Calibration Fixture Instruction Manual Model 280 286 Instruction Manual Part No 204210 Revision B ...

Page 2: ...l without first receiving the written permission of Fann Instrument Company Houston Texas U S A Printed in U S A NOTE Fann reserves the right to make improvements in design construction and appearance of our products without prior notice FANN is a registered trademark of Fann Instrument Company Fann Instrument Company P O Box 4350 HOUSTON TEXAS U S A 77210 TELEPHONE 281 871 4482 TOLL FREE 800 347 ...

Page 3: ...cedure Rheometer 3 3 Spring Calibration Procedure Model 280 Rheometer 7 4 Spring Calibration Procedure Model 286 Rheometer 11 5 Parts List 15 Table 1 Calibration Weight vs Dial Reading 4 Figure 1 Typical Torsion Shaft Assembly 5 2 Model 280 Set Up 6 3 Model 286 Set Up 10 ...

Page 4: ...ng the active length of the spring The Calibration Fixture may be used for Calibration of a replacement spring when it is being installed Verification of the calibration of the spring already in the instrument Re calibration of the spring already in the instrument Check the following parts of the Rheometer for runout or damage If any of these parts are bent or otherwise damaged repairs must be mad...

Page 5: ...2 ...

Page 6: ...e hole in the Pulley Support closest to the pulley for the Model 280 Rheometer and away from the pulley for the Model 286 Rheometer 4 Screw Calibration spool 5 onto bob shaft threads down until it tight Wind two to three turns of thread around the drum clockwise as viewed from the top of the spool Refer to Fig 2 for Model 280 and Fig 3 for Model 286 Route the thread over the sheave and allow it to...

Page 7: ...4 B Take readings as shown in the Table 1 TABLE 1 WEIGHT DIAL READING 25 GRAMS 103 104 50 GRAMS 204 206 75 GRAMS 309 312 ...

Page 8: ...5 Fig 1 Typical Torsion Shaft Assembly ...

Page 9: ...6 Fig 2 Model 280 Set Up ...

Page 10: ... straight up and remove 9 Loosen two Zeroing screws that act as set screws against the flat on the Torsion Body sufficiently to clear and allow removal of the torsion body 10 Remove the large retainer ring on top of the Torsion Body 11 Lift the torsion shaft assembly up and out of the Rheometer frame 12 Re assemble the Dial 60 onto the top of the Bob Shaft and secure it with screw 79 B Preliminary...

Page 11: ...s with the divisions on the dial and lock in place Loosen set screw 81 B to loosen collet 67 2 in the torsion body 61 then rotate the dial and bob shaft to align Zero with the hairline Lock collet 67 2 in this position with set screw 81 B 6 Screw Calibration spool 5 onto bob shaft threads down until it is tight Wind two to three turns of thread around the drum clockwise as viewed from the top of t...

Page 12: ...nst the flat on the torsion body to slightly rotate the torsion body as needed to align the 0 with the hairline Recheck using the cover and repeat if necessary Both zero adjust screws should be tight against the torsion body 8 Place cover in place and install the four 6 32 x 1 4 long cover to frame screws 9 Position shift arm in the STIR position then install the shift arm cam and bearing onto the...

Page 13: ...10 Fig 3 Model 286 Set Up ...

Page 14: ... the cover up and off Rheometer 9 Remove flat head screw 79 from top center of dial 60 then carefully lift dial straight up and remove 10 Remove printed circuit board by removing screws as follows Two 4 40 screws mounting the printed circuit board to the frame These two screws go through the circuit board from the top One is near each side of the board about 5 inches 12 7 cm from the square end On...

Page 15: ...sion shaft assembly Refer to Fig 3 1 Install the pulley 3 into the bottom hole of the pulley holder bracket 2 Loosen the two screws holding the lower bearing support bar 1 and position it at the bottom of the slots and re tighten the screws 3 Slide torsion shaft assembly into fixture from the top Position torsion body hub into the hole in the top of the fixture and bearing 68 in the hole in the lo...

Page 16: ... 61 and lock it with hex set screw 81 B in a position so that dial zero aligns with the hairline 5 Check calibration as described in Section 3D above If still incorrect repeat this section F Re assembly of the Torsion Shaft Assembly into the Rheometer 1 Loosen and rotate the hairline magnifier out of the way of removing the Torsion Shaft Assembly 2 Unscrew the Calibration spool 5 and the thread 6 ...

Page 17: ...to slightly rotate the torsion body as needed to align the 0 with the hairline Recheck using the cover and repeat if necessary Both zero adjust screws should be tight against the torsion body when adjustment is complete 10 Install Cover be installing the following fasteners Two 6 32 x 1 4 long cover to frame screws Two 6 32 x 1 4 long cover to power supply bracket screws Nut cover to speed selecto...

Page 18: ...grams 207849 Thread 204210 Instructions 205683 Bearing bob shaft B Optional Calibration Fluids PART NO DESCRIPTION 207124 Calibration Fluid 10 cP 207119 Calibration Fluid 20 cP 207120 Calibration Fluid 50 cP 207121 Calibration Fluid 100 cP 207122 Calibration Fluid 200 cP 207123 Calibration Fluid 500 cP ...

Page 19: ...inst possible damage or loss Fann will not be responsible for damage resulting from careless or insufficient packing Out of warranty products will be repaired for a nominal charge Return of Items Before returning items for any reason authorization must be obtained from Fann Instrument Company When applying for authorization please include information regarding the reason the items are to be return...

Reviews: