40i TS - v1205 - Installation/Operation - (22/29)
Important
Before capturing data for an accuracy graph, a home
search must be carried out because the compensation
will not be applied until the home search is done. To use
this compensation, it is recommended to force a home
search on power-up.
Pressing the Edit button displays a table with 105 points
and their corresponding error values.
Error to be compensated = Master's actual position -
displayed position
There is no need to use all the points. The compensation
point must have at least one point with no error (error 0).
Different options are offered when the Function button is pressed:
To exit the screen saving the data.
Draws a graph with the points and errors entered. It is
recommended to check the graph to detect possible mistakes
made when entering data.
Note:
The maximum slope for the compensation graph is ±0.8 mm/m
.
3.6.4
Test mode.
It may be used to know system information such as software version, hardware version, software
burning date, etc.
Pressing the Test key displays the software and hardware versions, software recording date,
checksum, error history, etc.
Pressing Test again offers the chance to run different tests that are very useful for troubleshooting
the DRO itself or the encoder.
The Test mode is for the installer only and the access is protected with the password:
231202
3.6.5
Machine Control
For setting the spindle control parameters, inputs and
outputs.
Edit
Exit
Function
Draw Graph
Function
Setup
Setup
Installer parameters
Machine Control