OPERATING INSTRUCTIONS
3-19
Figure 3-14
Probe Output versus Flux Angle
The probe is calibrated and specified with flux lines passing
perpendicularly through the Hall generator.
5) As shown in Figure 3-15 the greater the distance between the
magnetic source and the Hall probe the fewer flux lines will pass
through the probe, causing the probe’s output to decrease.
Figure 3-15
Probe Output versus Distance
6) Flux density can vary considerably across the pole face of a
permanent magnet. This can be caused by internal physical flaws