OPERATING INSTRUCTIONS
3-7
POWER-UP
Depress the POWER switch. There will be a momentary audible
beep and all display segments will appear on the display.
Figure 3-6
Power-Up Display
The instrument will conduct a self test before measurements
begin. If a problem is detected the phrase “Err” will appear on the
display followed by a 3-digit code. The circuitry that failed will be
retested and the error code will appear after each failure. This
process will continue indefinitely or until the circuitry passes the
test. A condition in which a circuit fails and then passes should
not be ignored because it indicates an intermittent problem that
should be corrected.
If the self test is successful the meter will perform a self
calibration. During this phase the meter will display the software
revision number, such as “r 1.0”. Calibration will halt if there is no
Hall probe connected. Until the probe is connected the phrase
“Err” will appear accompanied by a flashing “PROBE” annunciator
as shown in Figure 3-7.