![F.W. Bell 5060 Instruction Manual Download Page 34](http://html.mh-extra.com/html/f-w-bell/5060/5060_instruction-manual_535139034.webp)
OPERATING INSTRUCTIONS
3-17
Figure 3-13
Manual ZERO Function
By turning the MANUAL control in either direction the reading will
be altered. Turning the control clockwise adds to the reading,
turning it counterclockwise subtracts from the reading. Turning it
slowly results in a fine adjustment, turning it quickly results in a
coarse adjustment.
NOTE:
Making a manual ZERO adjustment not only affects the
lowest range but also the highest range, though to a lesser extent.
For example, assume an automatic ZERO has already been
performed, after which both ranges should read zero. Now a
manual adjustment is made that causes the reading on the lowest
range to be non-zero. The reading on the other range may also
Summary of Contents for 5060