background image

HI, LO and Deadband Setpoints 

<HI and LO Relay> 

Relays switch state when their user-programmable setpoint is exceeded. They switch again 

when the limit is no longer exceeded. If the measurement exceeds the high setpoint, the 

high relay will switch state. If the measurement is lower than the low setpoint, the low relay 

will switch state. The programming steps are detailed later in this section. 

 

<HI and LO Deadband> 

The Deadband feature is used to minimize relay chatter (rapid on/off relay switching) near 

setpoint. This is accomplished by creating a region where no relay action can occur. The 

High Deadband parameter creates a band at the high setpoint and the Low Deadband 

parameter creates a band at the low setpoint.   

 

HI relay ON: reading > (HI relay se HI deadband) 

HI relay OFF: reading < (HI relay setpoint) 

LO relay ON: reading < (LO relay setpoint - LO deadband) 

LO relay OFF: reading > (LO relay setpoint) 

 

 

 

HI setpoint 

pH 7 

LO setpoint 

HI deadband 

LO deadband 

HI relay ON 

LO relay ON 

Summary of Contents for 4801P

Page 1: ... mV 500 1999 mV 0 110 C Accuracy 0 01 1 digit 2 1 digit 0 2 1 digit Resolution 0 01 pH 0 1 1 mV 0 1 C Compensation ATC 0 100 C N A Deadband pH 0 00 to 2 00 pH ORP 0 to 200 mV Temperature Compensation ATC 0 100o C via 30K Thermistor or MTC Limit Relays 2 SPDT contacts rated for 5A at 115V Relay status indicators Front panel LED lamp Power supply 9V DC Panel Cutout 1 16th DIN 1 81 x 1 81 46 x 46mm M...

Page 2: ...l the bezel is flush with the panel 3 Slide the mounting bracket over the rear of the controller and press snugly against the rear of the panel Rear panel connection Connect the alarm control devices to the HI or LO relay terminals The SPDT relays are dry contact type and the terminal designations are C Common NO Normally Open NC Normally Closed ...

Page 3: ...ensation ATC or Manual temperature compensation MTC 7 Temperature reading and unit 8 unit of pH or ORP reading 1 Turn on off the controller 2 Enter calibration mode 1 Switch pH and ORP mode 2 Press and hold to enter setting mode 3 In setting mode press to store each setting 1 In measurement mode adjust temperature value of MTC 2 In setting mode adjust each parameter 1 2 3 4 8 7 6 5 ...

Page 4: ...rode must be replaced A slope of 100 is ideal Note 1 Calibration error indicator icon will appear and Err instead of SA if calibration fails 2 When doing a 2 or 3 points calibration Calibrate with buffer pH 7 first and then follow with buffer pH 4 or pH 10 pH Calibration type and temperature unit setting This controller has two pH calibration type USA and NIST and the calibration points are follow...

Page 5: ...ection HI and LO Deadband The Deadband feature is used to minimize relay chatter rapid on off relay switching near setpoint This is accomplished by creating a region where no relay action can occur The High Deadband parameter creates a band at the high setpoint and the Low Deadband parameter creates a band at the low setpoint HI relay ON reading HI relay setpoint HI deadband HI relay OFF reading H...

Page 6: ...each setting 3 The display will then automatically cycle through 5 settings rl relay on or off HIr HI relay setpoint Lor LO relay setpoint dbH HI Deadband and dbL LO Deadband 4 When setting HI or LO setpoints press and hold to switch the value to Positive or to Negative Choose pH or ORP Enter setting mode HIr HI relay setpoint Lor LO relay setpoint dbH HI Deadband dbL LO Deadband End ON OFF Relay ...

Page 7: ...ht is always on OFF back light is always off Auto back light will be on for 30 seconds while any button is pressed and then off Reset the controller to default setting 1 Press and hold until a beep sound release and then press and hold again until a beep sound 2 Press to switch OFF to On and then press to reset the controller to default setting ...

Reviews: