3. On the touch screen, press [CCdual] to change it into Dual Cross Cylinder Mode.
4. From the patient’s view, the Dot-Group Chart is separately
shown divided as two.
▶
If the left one is seen clearly, turn the dial counterclockwise (+ direction.)
▶
If the right one is seen clearly, turn the dial clockwise (- direction.)
▶
If the patient says there is no differene, finish the test.
5. Press [A] or left/right button to estimate the distorted vision axis.
Press [C] or left/right button to estimate the circumference power.
[Figure 10-13] Dual Cross Cylinder Test (Ditorted Vision Axis Test)
Digital Refractor EDR-9000