9.1.2
Cro s s Cylin d e r Te s t us in g Do ts Cha rt ( AXIS )
The 2
nd
step of the System-Defined Program is to revise the value of AXIS for the right eye more
elaborately, based on the result of 1st step SPH-CYL-AXIS test. In case of Jackson Cross Cylinder
Test, ask the patient with which eye he looks more clearly, while pressing number [1] and [2] buttons
alternately. When he says number [1], turn the dial in + direction (countercloskwise,) while number
[2] in – direction (clockwise.) Continue this until the patient answers that he sees both with similar
clarity.
[Figure 9-3] 2
nd
step Cross Cylinder Test (Axis)
9.1.3
Cro s s Cylin d e r Te s t us in g Do ts Cha rt ( CYL)
The 3
rd
step of the System-Defined Program is to revise the value of CYL for the right eye more
elaborately, based on the result of 1st step SPH-CYL-AXIS test. In case of Jackson Cross Cylinder
Test, ask the patient with which eye he looks more clearly, while pressing number [1] and [2] buttons
alternately. When he says number [1], turn the dial in + direction (countercloskwise,) while number
[2] in – direction (clockwise.) Continue this until the patient answers that he sees both with similar
clarity.
[Figure 9-4] 3rd step Cross Cylinder Test (CYL)
Digital Refractor EDR-9000
Summary of Contents for EDR 9000
Page 1: ... EDR 9000 Digital Retractor Operation Manual ED R 9000 Digital Retractor ...
Page 2: ......
Page 11: ...9 Appendix Hardware Specification 115 Digital Refractor EDR 9000 ...
Page 64: ...62 Operation Operation Manual Manual Figure 9 2 Adjustment of 1 st step S C A Value ...
Page 71: ...69 Figure 9 14 Far Distance Vertiacl Coincidence Aniseikonia Test Digital Refractor EDR 9000 ...