background image

1. 

Check all bolts to ensure they are tight. 

Remove covers as necessary for inspection.

2. 

Check arm to mast pivot bolt 

by removing plastic cap to assure nut is tight and cotter pin  

     is in place. If plastic cap is missing order replacement.

3. 

Foot platform assembly pins need to checked to assure  

    they have not pushed in and still engage into base.

 
4.

 Check bolts on back of shin pad.

4.

 Safety tabs need to be checked 

to make sure they are installed correctly, not missing or torn.

 

PIN ON FOOT
PLATFORM

ARM TO MAST

BASE                          PIVOT                    REAR WHEEL           FRONT WHEEL               LINKAGE                  SPREADER            

EZ Way Smart Stand

®

 

Safety

 & 

Maintenance Checklist

EZ Way, Inc.

The EZ Way Smart Stand

®

 requires a minimum of servicing to keep it in good working order.  Nevertheless, it is  

important that certain basic checks be periodically made by maintenance staff to ensure on-going safety throughout the 
life of the device.

The manufacturer suggests that the following components and operating points be scheduled for inspection at intervals 
not greater than six months.  Any detected deficiency must be rectified before the stand is put back into service.

1

2

3

4

5

Summary of Contents for Smart Stand S800 series

Page 1: ...Service Manual Model S800 Smart Stand Form 1 146 Rev 2 5 19 WARNING Cancer and Reproductive Harm www P65Warnings ca gov...

Page 2: ...ax 712 542 1899 www ezlifts com Table of Contents Parts Breakdown 3 Monthly Maintenance Checklist 7 Smart Stand Operating Instructions 9 Scale Calibration 10 Advanced Smart Stand Features 11 Wiring Sc...

Page 3: ...EZ Way Inc EZ Way Inc 800 627 8940 Fax 712 542 1899 www ezlifts com 3 S800PN Non Scale Parts Diagram...

Page 4: ...ck Assembly UL 6 12743 Cover Base Assembly Stand 800 lb 7 11885 Caster Rear Lift 1000 lb Stand 800 lb 8 13310 Bar Pivot Chrome Plated Lift Stand 9 12745 Platform Foot Assembly Stand 800 lb 10 14500 Ca...

Page 5: ...EZ Way Inc EZ Way Inc 800 627 8940 Fax 712 542 1899 www ezlifts com 5 S800PS With Scale Parts Diagram...

Page 6: ...11871 Battery Pack Assembly UL 8 12743 Cover Base Assembly Stand 800 lb 9 11885 Caster rear Lift 1000 lb Stand 800 lb 10 13310 Bar Pivot Chrome Plated Lift Stand 11 12745 Platform Foot Assembly Stand...

Page 7: ...not missing or torn PIN ON FOOT PLATFORM ARM TO MAST BASE PIVOT REAR WHEEL FRONT WHEEL LINKAGE SPREADER EZ Way Smart Stand Safety Maintenance Checklist EZ Way Inc The EZ Way Smart Stand requires a mi...

Page 8: ...Test the Emergency Stop Switch If the switch does not stop the stand when activated call the EZ Way Service Department immediately Perform a functional test of the emergency lowering feature of the s...

Page 9: ...a symbol indicating that the weight is recalled MENUS Enter Customer Settings Menu Weight Calibration 1 After power up sequence press and hold the Down key followed quickly by the Up key and hold bot...

Page 10: ...to ensure the scale is functional 5 Enter the calibration menu After power up sequence press and hold the DOWN key followed quickly by the UP key and hold both keys for 5 seconds 6 Press the UP key to...

Page 11: ...ustomer will need to wait 15 minutes to cool the actuator motor Actuator Overload For Arm Protection 1 When an Up or Down command is executed the circuit board monitors current consumed by the actuato...

Page 12: ...me since the last inspection exceeds 30 days a reminder message will appear when the unit is powered up The purpose of the inspection reminder is to notify the customer that the lift or stand needs to...

Page 13: ...s read only information provides customer feedback for accidental lifting of a trapped harness accidental lifting of arm against obstruction or accidental lifting of overweight patient Lift Stand Tota...

Page 14: ...EZ Way Inc EZ Way Inc 800 627 8940 Fax 712 542 1899 www ezlifts com 14 Smart Stand Wiring Schematic...

Page 15: ...ake sure on off button is not flattened out Check if green LED is flashing on board Flashing LED indicates the PCB is getting power Make sure W6 jumper is in place No up down but display works Check c...