Advanced Smart Stand

®

 Features

ADVANCED DIAGNOSTICS

Real Time Current – For Actuator Diagnostics

1.  When the display shows “EZ Stand Awake”, press the hidden key.  Now whenever the stand is in 

motion, the real time current will be shown.

2.  The Real Time Current function will disable after power cycling the unit.

Real Time Voltage – For Battery Diagnostics

1.  In the case of a possible battery or connection issue the battery voltage can be monitored real-time. 

Go to the Customer Settings menu.

2.  Press the Enter key several times until you see “Volts = …”.

CUSTOMER SETTINGS MENU (FULL DESCRIPTIONS)

Within the Customer Settings Menu there are several functions that provide customer feedback. 
The functions include; an Inspection Reminder, stand usage, and diagnostic information. Below is a 
description of each function within the Customer Settings Menu.

Inspection Reminder – For Scheduled Maintenance

The Inspection Reminder function tracks the last inspection occurrence. When the duration of time 
since the last inspection exceeds 30 days a reminder message will appear when the unit is powered 
up. The purpose of the inspection reminder is to notify the customer that the lift or stand needs to be 
inspected with the ‘EZ Way Safety & Maintenance Checklist’ form (see attached). When the checklist is 
complete the reminder is to be reset in the Customer Settings Menu as described below.
1.  To enable/disable the Inspection Reminder enter the Customer Settings Menu.
2.  Press enter. Scroll to Inspection Reminder. Press enter. Scroll to choose Yes or No. Press Enter.
3.  Press Exit twice.

Inspection Complete

1.  If ‘INSPECTION OVERDUE’ appears during power up complete the ‘EZ Way Safety & 

Maintenance Checklist’.

2.  Enter the Customer Settings Menu.
3.  Press enter. Scroll to Inspection Complete. Press enter. Scroll to choose Yes. Press Enter.
4.  Press Exit twice.

Overweight Counts – For Diagnostic Tracking with Scale  

The Overweight Counts function tracks ‘Weight Overload’ occurrences. When a ‘WEIGHT 
OVERLOAD’ occurs the ‘Overweight Counts’ register increments one. This read-only information 
provides customer feedback for; accidental lifting of a trapped harness, or accidental lifting of overweight 
patient. This information is available for units with the scale option only.

Summary of Contents for Smart Stand S800 series

Page 1: ...Service Manual Model S800 Smart Stand Form 1 146 Rev 2 5 19 WARNING Cancer and Reproductive Harm www P65Warnings ca gov...

Page 2: ...ax 712 542 1899 www ezlifts com Table of Contents Parts Breakdown 3 Monthly Maintenance Checklist 7 Smart Stand Operating Instructions 9 Scale Calibration 10 Advanced Smart Stand Features 11 Wiring Sc...

Page 3: ...EZ Way Inc EZ Way Inc 800 627 8940 Fax 712 542 1899 www ezlifts com 3 S800PN Non Scale Parts Diagram...

Page 4: ...ck Assembly UL 6 12743 Cover Base Assembly Stand 800 lb 7 11885 Caster Rear Lift 1000 lb Stand 800 lb 8 13310 Bar Pivot Chrome Plated Lift Stand 9 12745 Platform Foot Assembly Stand 800 lb 10 14500 Ca...

Page 5: ...EZ Way Inc EZ Way Inc 800 627 8940 Fax 712 542 1899 www ezlifts com 5 S800PS With Scale Parts Diagram...

Page 6: ...11871 Battery Pack Assembly UL 8 12743 Cover Base Assembly Stand 800 lb 9 11885 Caster rear Lift 1000 lb Stand 800 lb 10 13310 Bar Pivot Chrome Plated Lift Stand 11 12745 Platform Foot Assembly Stand...

Page 7: ...not missing or torn PIN ON FOOT PLATFORM ARM TO MAST BASE PIVOT REAR WHEEL FRONT WHEEL LINKAGE SPREADER EZ Way Smart Stand Safety Maintenance Checklist EZ Way Inc The EZ Way Smart Stand requires a mi...

Page 8: ...Test the Emergency Stop Switch If the switch does not stop the stand when activated call the EZ Way Service Department immediately Perform a functional test of the emergency lowering feature of the s...

Page 9: ...a symbol indicating that the weight is recalled MENUS Enter Customer Settings Menu Weight Calibration 1 After power up sequence press and hold the Down key followed quickly by the Up key and hold bot...

Page 10: ...to ensure the scale is functional 5 Enter the calibration menu After power up sequence press and hold the DOWN key followed quickly by the UP key and hold both keys for 5 seconds 6 Press the UP key to...

Page 11: ...ustomer will need to wait 15 minutes to cool the actuator motor Actuator Overload For Arm Protection 1 When an Up or Down command is executed the circuit board monitors current consumed by the actuato...

Page 12: ...me since the last inspection exceeds 30 days a reminder message will appear when the unit is powered up The purpose of the inspection reminder is to notify the customer that the lift or stand needs to...

Page 13: ...s read only information provides customer feedback for accidental lifting of a trapped harness accidental lifting of arm against obstruction or accidental lifting of overweight patient Lift Stand Tota...

Page 14: ...EZ Way Inc EZ Way Inc 800 627 8940 Fax 712 542 1899 www ezlifts com 14 Smart Stand Wiring Schematic...

Page 15: ...ake sure on off button is not flattened out Check if green LED is flashing on board Flashing LED indicates the PCB is getting power Make sure W6 jumper is in place No up down but display works Check c...

Reviews: