8
Base Setting
A
B
B
Transporting a patient without an installed
brace is not allowed because of safety reasons.
After the toilet is used, the brace absolutely
must be re-installed!
Step 1
Before disassembling brace (A), lock both rear
wheels with both wheel locks (B) (see: 8.9).
Caution – Safety Advisory!
Step 2
Pull out the 2 clips (C) from the brace.
C
C
Step 3
Push back the brace to
the striker and put one
clip back in. On the other
side, insert the clip in the
swivel arm (D).
D
Step 4
Swing out both swivel arms of the brace toward the
inside of the chair so that it can be rolled over the
toilet. Release the wheel brake locks.
Step 5
After toilet use is finished, lock the two rear wheels
once again and install the stabilization brace
following the process in reverse.
Use the chair only when it is in the
base setting.
9. Troubleshooting
Fault
Possible Cause
Correction
Armrests do
not swivel
properly
Locking pins
are caked with
deposits
Clean, check
periodically
Locking pins
cannot be
pulled out
Caked up,
tension spring
is jammed or
defective
Clean,
replace
locking pin
Footrests do
not swivel
properly
Caked up
joints,
jamming
Clean,
check screw
connection
Castor wheels
do not lock
Generally
defective
See Section
Maintenance
Wheel turns
with difficulty
Deposit
build-up
Clean
Wheel does
not rotate
Generally
defective
See section on
Maintenance
Wheel do
not Rotate
Generally
defective
See section on
Maintenance
Wheels no
longer round
Wheels
unevenly worn
See section on
Maintenance
Chair frame
too unstable
Screws are
loose
Tighten frame
and castor screws
Castors are
bent
Bumped into
objects
See section on
Maintenance
End Setting