(1) Make sure that drain stopcock is attached to drain hose.
(2) Fill the water (approx. 14
ℓ
) to soak cooling coil into low
temperature circulation bath.
※
Please put a water in the circulation bath slowly.
3. Filling cool water
Water level of the cooling
water circulator bath
(3) Make sure that water-out valve for
adjusting flow
volume is closed.
Make sure that power switch is turned off
and plug the mains plug into AC outlet.
4. Connecting mains plug
(4) Set the bath cover on the main unit.
*
Use antifreeze if you use the unit at
+
7
℃
or
lower. However, if you use ethylene glycol or
Nybrine, the viscosity will be higher in low
temperature, which makes temperature
distribution in the bath worse.
In such a case, mix the moderate amount of
water ( Make sure the freezing temperature
when the liquid is concentrated.)
*
Make sure that there’s no foreign substance in
circulation liquid, which will cause malfunction.
Do not use pure water
Use tap water or softened tap water.
Do not use pure water or ion exchanged water.
These types of water solve carbon dioxide in
the air into acid solution, which could
corrode the metal of circulation route.
Also, pin hole of cooling coil or circulation
pump may be impaired because of the use of
these water.
14
Valve for adjusting flow volume
Close
Open
WATER IN
WATER OUT
※
Do not move the unit ,
after filling liquid in the bath.