background image

VB400-EU-EN-V1.0 

Zero Adjustment Procedure 

Over time the meter ‘0” may drift by a few digits. This is caused by temperature change, battery 

voltage and other ageing factors. Usually this small change is not of significance for typical 

measurements. To remove the error: 

1. 

Open the battery compartment 

2. 

Hold the battery lid down so contact is made and the meter can be turned on. 

3. 

With no vibration, adjust the meter zero adjustment until the display reads zero (no minus 

sign). 

 

 

 

 

 

 

Battery lid (1) and Zero adjustment (2) 

Battery Replacement 

Unscrew the meter’s top cover 

Lift the battery lid and replace the four AAA batteries, observing polarity 

Replace the cover. 

ISO Vibration Severity Guidelines 

ISO 10816 

Standards provide guidance for evaluating vibration severity in machines operating in 

the 10 to 200Hz (600 to 12,000 RPM) frequency range. Examples of these types of machines are 
small, direct-coupled, electric motors and pumps, production motors, medium motors, generators, 
steam and gas turbines, turbo-compressors, turbo-pumps and fans. The axis of the rotating shaft 
may be horizontal, vertical or inclined at any angle.  

 

 

1

2

Summary of Contents for VB400

Page 1: ...User s Guide Pen Vibration Meter Model VB400 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1 800 517 8431 Visit us at www TestEquipmentDepot com ...

Page 2: ...ce Features Applications for industrial vibration monitoring All in one pen type digital vibration meter Acceleration Velocity measurement RMS measurement value Metric Imperial display unit Frequency range 10 Hz 1 kHz sensitivity designed to meet ISO 2954 High accuracy and easy to read LCD display Complete with the test pin and magnetic base Low battery indicator IP65 protection Safety This symbol...

Page 3: ...3 Hold Button 4 Function Button 5 Battery Cover Compartment 6 Vibration Sensor 7 Sensing Head 8 Magnetic Base Display Icons HOLD HOLD has been activated ACC Acceleration with units m s 2 ft s 2 g VEL Velocity with units mm s cm s inch s Low battery 1 2 3 4 5 6 7 8 ...

Page 4: ...utton to step through and set the function velocity or acceleration and the units desired 2 For non ferrous surfaces gently hold the meter as shown with the sensing head against the vibrating surface 3 For ferrous surfaces attach the meter to the surface with the magnetic mount 4 The meter must be held perpendicular to the surface to maintain measurement accuracy 5 Read the measured value in the d...

Page 5: ...no minus sign Battery lid 1 and Zero adjustment 2 Battery Replacement 1 Unscrew the meter s top cover 2 Lift the battery lid and replace the four AAA batteries observing polarity 3 Replace the cover ISO Vibration Severity Guidelines ISO 10816 Standards provide guidance for evaluating vibration severity in machines operating in the 10 to 200Hz 600 to 12 000 RPM frequency range Examples of these typ...

Page 6: ...40 x 32 mm 6 9 x 1 6 x 1 3 inch Sensing head Round 9 mm Dia x 30 mm Accuracy Specifications Mode Range Resolution Accuracy 0 5 to 199 9m s 2 0 1 0 05 to 20 39g 0 01 2 to 656ft s 2 1 5 rdg 2digits 160Hz 80Hz 23 5 C Acceleration Calibration point 50m s 2 160Hz 0 5 to 199 9 mm s 0 1mm s 0 05 to 19 99 cm s 0 01 0 02 to 7 87inch s 0 01inch s 5 rdg 2digits 160Hz 80Hz 23 5 C Velocity Calibration point 50...