UM200-en-GB_V1.5   2/17 

Recalling pre-stored 

/LENGTH data 

There are up to 20 pre-stored Resistance per unit values stored into memory. These values can be 
recalled and used for cable length measurement. 

1. 

In SETUP mode, press the 

RECALL

 button to recall pre-stored data.  

2. Press 

the 

SELECT

 button until the resistance per unit values appear. 

3. Press 

the 

RECAL

L button to step through the stored values. 

4. 

Turn the rotary switch to the proper range and press 

START

 to measure the cable length. 

NOTE: 

The range of length is 0.0001 ft or m to 9999K ft or m. 
When the cable is disconnected, OL 

 is displayed instead of OL FT or m.  

When the resistance is 0, 0 

 is shown instead of 0 FT or m. 

If the length is less than 0.0001 feet or meters but greater than 0, 0.0001 feet or meters is shown. 

 

Memory Record, Read and Clear 

Record 

1. 

With a reading in the display, press the 

REC / READ

 button. 

2. 

The memory location number will flash on the display and the data will be stored to that 
location. 

Read 

1. 

Turn the rotary switch to SETUP. 

2. Press 

the 

REC / READ

 button. The memory will flash and then the data in that location will 

be displayed. 

3. Press 

the 

▲▼

 buttons to scroll through stored data. 

Clear 

1. 

Turn the meter off. 

2. 

Press and Hold the 

REC / READ

 button and simultaneously turn the meter on. 

NOTE: 

The memory capacity is 3000 data records.  When 3000 records are exceeded, there will be a long 
beeping sound and no further data can be recorded..

 

Backlight 

Press the 

 button to turn the backlight on or off. 

 

Summary of Contents for UM200

Page 1: ...User Manual Micro Ohmmeter Model UM200 Additional User Manual Translations available at www extech com ...

Page 2: ... ohmmeter Do not use the lithium battery to power anything but this instrument Do not soak or immerge the lithium battery in any liquid The lithium battery may explode if disposed of in Fire Recycled or disposed of the battery as directed by local regulations Do not expose the lithium battery to temperature higher than 60 C or 140 F Do not disassemble the lithium battery assembly Do not short circ...

Page 3: ...measurements Press the ZERO RECALL button to enter Relative mode the present readings will be reset to zero or press the ZERO RECALL button for over 2 seconds to exit the Relative mode In SETUP mode use this button to recall the pre stored sets of HI limits LO limits and resistance per unit 6 START STOP button In resistance mode press the START STOP button to start measurement Press the START STOP...

Page 4: ...ton to enable disable the backlight 14 button for 600mΩ 1A and 60mΩ 10A ranges Press to measure inductive materials when the LED is off 15 button Press the Beep button to turn on off the alarm function HI and LO When the alarm function is enabled the icon will be shown in LCD If the value of resistance is within the HI LO range a PASS will be displayed If the value of resistance is out of HI LO ra...

Page 5: ...r Clip Terminal or 4 wire Test Leads Terminal 2 P1 Alligator Clip Terminal or 4 wire Test Leads Terminal 3 P2 Alligator Clip Terminal or 4 wire Test Leads Terminal 4 C2 Alligator Clip Terminal or 4 wire Test Leads Terminal 5 Kelvin Clip Terminal 6 Kelvin Clip Terminal 7 POWER for the input of AC adaptor 1 2 3 4 1 2 3 4 5 6 7 ...

Page 6: ...light with inductance symbol is on this indicates the measurement is for both resistive materials and inductive materials When the LED light with inductance symbol is off inductive material cannot be measured NOTE The battery will not charge with the rotary switch in the OFF position Place the rotary switch in any range position to charge the battery Warning Do not measure any resistance object wi...

Page 7: ...together and press the Zero button The reading should read zero Ohms 4 Connect the test leads to the device under test 5 Press the START STOP button to start the measurement 6 The LCD will continuously display the value of the resistance Only resistive materials can be measured 7 If you need to change ranges Zero the range before making a measurement 8 To stop the measurement press the START STOP ...

Page 8: ... x10 button to move the decimal to the next position 7 To store the modified data the SELECT button NOTE If the reading of resistance is OL the unit will not indicate PASS or beep The alarm function works only when the resistance is within measuring range NOTE The limits of HI and LO alarms are 0 001mΩ and 999 99Ω The limits of Resistance per unit are 0 001μΩ ft or m 999 99Ω m 304 79Ω ft Cable Len...

Page 9: ...nce is 0 0 Ω is shown instead of 0 FT or m If the length is less than 0 0001 feet or meters but greater than 0 0 0001 feet or meters is shown Memory Record Read and Clear Record 1 With a reading in the display press the REC READ button 2 The memory location number will flash on the display and the data will be stored to that location Read 1 Turn the rotary switch to SETUP 2 Press the REC READ butt...

Page 10: ...battery part BATT 111V Never dispose of used batteries or rechargeable batteries in household waste As consumers users are legally required to take used batteries to appropriate collection sites the retail store where the batteries were purchased or wherever batteries are sold Disposal Do not dispose of this instrument in household waste The user is obligated to take end of life devices to a desig...

Page 11: ...000mΩ 5 000mΩ to 60 000mΩ 1 A 4 00mΩ to 40 00mΩ 10μΩ 0 25 250μΩ 15 00mΩ to 160 00mΩ 50 00mΩ to 600 00mΩ 100 mA 0 0400Ω to 0 4000Ω 100μΩ 0 25 2 5mΩ 0 1500Ω to 1 6000Ω 0 5000Ω to 6 0000Ω 10 mA 0 400Ω to 4 000Ω 1 mΩ 0 25 25mΩ 1 500Ω to 16 000Ω 5 000Ω to 60 000Ω 1 mA 4 00Ω to 40 00Ω 10 mΩ 0 25 250mΩ 15 00Ω to 160 00Ω 50 00Ω to 600 00Ω 100μA 0 0400kΩ to 0 4000kΩ 100 mΩ 0 75 3Ω 0 1500kΩ to 1 6000kΩ 0 50...

Reviews: