Alarm Function
Press the
button to enable the alarm function after the HI and LOW limits are set. The
symbol will be displayed in the LCD.
If the resistance measured falls within the range of HI and LO limits, a symbol of PASS is shown
in the LCD. If not, the buzzer will beep to indicate a fail.
If the reading is OL, the alarm function is temporarily disabled until a reading is obtained.
Alarm HI, LO, or Resistance per unit Setup
1.
Turn the rotary switch to
SETUP
.
2. Press
the
SELECT
button to select HI limit, LO limit or Resistance per unit
3. Press
the
ZERO / RECALL
button to step through existing limits (20 available) or to set a new
limit.
4. Press
the
SELECT
button to select a displayed limit
5.
To set a new limit, press the
▲
or
▼
button to increment or decrement the displayed value. To
increment or decrement value faster, press and hold the
▲
or
▼
button for over 2 seconds.
6. Press
the
RANGE x10
button to move the decimal to the next position.
7.
To store the modified data, the
SELECT
button.
NOTE:
If the reading of resistance is “OL”, the unit will not indicate PASS or beep.
The alarm function works only when the resistance is within measuring range.
NOTE:
The limits of HI and LO alarms are 0.001m
Ω
and 999.99
Ω
The limits of Resistance per unit are 0.001
μΩ
/ft (or m), 999.99
Ω
/m, 304.79
Ω
/ft
Cable Length Measurement
1. Prepare a 1 foot or a 1-meter-long sample from the cable to be measured.
2. Select the proper range of resistance, press the
START
button and measure the resistance of
the sample.
3. Press
the
Ω
/LENGTH
button for over 2 seconds. The meter will beep and the LCD will display
units of FT or m. The measured value is stored and used to determine cable length.
4. If needed, press the
FEET/METERS
button to change the units. If the units are changed, press
the
Ω
/LENGTH
button for over 2 seconds again.
5. Disconnect the sample cable and connect the Kelvin clips to the cable to be measured. The
LCD will display the length of the cable.
6. If “OL” appears in the display, select a higher range and press
START.