Specifications

 

Display

 

LCD

 

size:

 

41mm

 

x

 

34mm

 

(1.6

 

x

 

1.3”)

 

 

Range

 

0.00

 

to

 

49.99

 

NTU

 

(Nephelometric

 

Turbidity

 

Unit)

 

 

50

 

to

 

1000

 

NTU

  

Resolution

 

0.01

 

NTU/1

 

NTU

 

Accuracy

 

Range

 

0

 

to

  

49.99:

    

±5%

 

reading

 

or

 

±

 

0.5

 

NTU

 

whichever

 

is

 

greater

 

Range

 

50

 

to

 

1000:

    

±5%

 

reading

 

Operating

 

Temperature

  

0

 

to

 

50

o

C

 

(32

 

to

 

122

o

F)

 

Operating

 

Humidity

 

Less

 

than

 

85%

 

RH

 

Light

 

Source

  

LED,

 

850

 

nm

 

Light

 

Detector

 

Photo

 

Diode

 

Response

 

Time

 

Less

 

than

 

10

 

seconds

 

Sample

 

Volume

 

10mL

 

minimum

 

Power

 

Six

 

1.5V

 

AAA

 

batteries

  

Power

 

consumption

 

Standby

 

Mode:

 

approx.

 

3.5mADC;

 

Testing

 

Mode:

 

approx.

 

36mADC

 

Dimensions/Weight

 

155

 

x

 

76

 

x

 

62mm

 

(6.1

 

x

 

3.0

 

x

 

2.4”)

 

;

 

320g

 

(0.70lbs)

 

Auto

 

Power

 

OFF

 

Automatically

 

switches

 

off

 

10

 

minutes

 

after

 

the

 

last

 

key

 

press

 

Standard

 

Meets

 

ISO

 

7027

 

 

 

Copyright

 

©

 

2013

2016

 

FLIR

 

Systems,

 

Inc.

 

All

 

rights

 

reserved

 

including

 

the

 

right

 

of

 

reproduction

 

in

 

whole

 

or

 

in

 

part

 

in

 

any

 

form

 

ISO

9001

 

Certified

 

www.extech.com

 

 

 

 

www.

.com

[email protected]

1.800.561.8187

Summary of Contents for TB400

Page 1: ...USER GUIDE Turbidity Meter Model TB400 www com information itm com 1 800 561 8187 ...

Page 2: ...liable service Please visit our website www extech com to check for the latest version of this User Guide Product Updates and Customer Support Applications Typical applications include the measurement of municipal water food and beverage water or other aqueous solutions where fluid clarity is important Meter Description 1 LCD Display 2 HOLD button 3 REC MAX MIN button 4 POWER button 5 ZERO button ...

Page 3: ... display will flash tESt for approximately 10 seconds and then the turbidity value in NTU units will be displayed 7 Momentarily press the OFF button to turn off the meter Data HOLD Press the HOLD button once while taking a measurement to freeze the value on the display Press HOLD again to release the Data Hold function Data Record MAX MIN Readings 1 The data record function records the MAX and Min...

Page 4: ...d close and latch the cover Turn on the meter and Press and hold the CAL button until the LCD displays CAL then release the CAL button 2 Wait one second and then the LCD will display Note If the LCD does not display zero when using 0 NTU liquid press the ZERO button continuously until the LCD displays a zero value 3 The meter is now ready for 0 NTU calibration 4 Momentarily press CAL button The LC...

Page 5: ...ntarily press the CAL button The LCD will flash CAL for 10 seconds 9 When calibration is complete the LCD will return to normal operating mode NOTE If the following error message appears Error cannot be calibrated check the value of the standard solution If error message still appears complete the Calibration Clear procedure CAL 100 ntu www com information itm com 1 800 561 8187 ...

Page 6: ...on continuously until the LCD displays CAL 4 Wait one second the LCD will then display 5 Momentarily press the HOLD button The LCD will display 6 Momentarily press the HOLD button again The LCD will display 7 Momentarily Press the CAL button The LCD will flash CAL for 10 seconds 8 The calibration data clear is complete The meter will return to normal operating mode CAL ntu 0 00 CAL ntu CLr CAL ntu...

Page 7: ...s may explode or leak Never mix battery types Always install new batteries of the same type Never dispose of used batteries or rechargeable batteries in household waste As consumers users are legally required to take used batteries to appropriate collection sites the retail store where the batteries were purchased or wherever batteries are sold Disposal Do not dispose of this instrument in househo...

Page 8: ... 850 nm Light Detector Photo Diode Response Time Less than 10 seconds Sample Volume 10mL minimum Power Six 1 5V AAA batteries Power consumption Standby Mode approx 3 5mADC Testing Mode approx 36mADC Dimensions Weight 155 x 76 x 62mm 6 1 x 3 0 x 2 4 320g 0 70lbs Auto Power OFF Automatically switches off 10 minutes after the last key press Standard Meets ISO 7027 Copyright 2013 2016 FLIR Systems Inc...

Reviews: