background image

TB400-en-GB_V1.3   8/16 

5

7.

 

Gently

 

shake

 

the

 

100

 

NTU

 

test

 

solution

 

bottle,

 

Line

 

up

 

the

 

white

 

mark

 

on

 

the

 

bottle

 

and

 

insert

 

it

 

into

 

the

 

test

 

chamber

 

and

 

close

 

and

 

lock

 

the

 

cover.

 

8.

 

Momentarily

 

press

 

the

 

"CAL"

 

button.

 

The

 

LCD

 

will

 

flash

 

CAL

 

for

 

10

 

seconds.

 

 

 

 

 

9.

 

When

 

calibration

 

is

 

complete

 

the

 

LCD

 

will

 

return

 

to

 

normal

 

operating

 

mode.

 

 

NOTE:

 

If

 

the

 

following

 

error

 

message

 

appears

 

"Error

 

cannot

 

be

 

calibrated..."

 

check

 

the

 

value

 

of

 

the

 

standard

 

solution.

 

If

 

error

 

message

 

still

 

appears

 

complete

 

the

 

"Calibration

 

Clear"

 

procedure.

 

 

CAL 

 

100 

     ntu 

www.

.com

[email protected]

1.800.561.8187

Summary of Contents for TB400

Page 1: ...USER GUIDE Turbidity Meter Model TB400 www com information itm com 1 800 561 8187 ...

Page 2: ...liable service Please visit our website www extech com to check for the latest version of this User Guide Product Updates and Customer Support Applications Typical applications include the measurement of municipal water food and beverage water or other aqueous solutions where fluid clarity is important Meter Description 1 LCD Display 2 HOLD button 3 REC MAX MIN button 4 POWER button 5 ZERO button ...

Page 3: ... display will flash tESt for approximately 10 seconds and then the turbidity value in NTU units will be displayed 7 Momentarily press the OFF button to turn off the meter Data HOLD Press the HOLD button once while taking a measurement to freeze the value on the display Press HOLD again to release the Data Hold function Data Record MAX MIN Readings 1 The data record function records the MAX and Min...

Page 4: ...d close and latch the cover Turn on the meter and Press and hold the CAL button until the LCD displays CAL then release the CAL button 2 Wait one second and then the LCD will display Note If the LCD does not display zero when using 0 NTU liquid press the ZERO button continuously until the LCD displays a zero value 3 The meter is now ready for 0 NTU calibration 4 Momentarily press CAL button The LC...

Page 5: ...ntarily press the CAL button The LCD will flash CAL for 10 seconds 9 When calibration is complete the LCD will return to normal operating mode NOTE If the following error message appears Error cannot be calibrated check the value of the standard solution If error message still appears complete the Calibration Clear procedure CAL 100 ntu www com information itm com 1 800 561 8187 ...

Page 6: ...on continuously until the LCD displays CAL 4 Wait one second the LCD will then display 5 Momentarily press the HOLD button The LCD will display 6 Momentarily press the HOLD button again The LCD will display 7 Momentarily Press the CAL button The LCD will flash CAL for 10 seconds 8 The calibration data clear is complete The meter will return to normal operating mode CAL ntu 0 00 CAL ntu CLr CAL ntu...

Page 7: ...s may explode or leak Never mix battery types Always install new batteries of the same type Never dispose of used batteries or rechargeable batteries in household waste As consumers users are legally required to take used batteries to appropriate collection sites the retail store where the batteries were purchased or wherever batteries are sold Disposal Do not dispose of this instrument in househo...

Page 8: ... 850 nm Light Detector Photo Diode Response Time Less than 10 seconds Sample Volume 10mL minimum Power Six 1 5V AAA batteries Power consumption Standby Mode approx 3 5mADC Testing Mode approx 36mADC Dimensions Weight 155 x 76 x 62mm 6 1 x 3 0 x 2 4 320g 0 70lbs Auto Power OFF Automatically switches off 10 minutes after the last key press Standard Meets ISO 7027 Copyright 2013 2016 FLIR Systems Inc...

Reviews: