Setting the Datalogger Sampling Time (Rate) ‘SP-t’
1.
From the
SP-t
parameter the sampling rate can be set to ‘0’ seconds (for manual logging) or
1, 2, 5, 10, 30, 60, 120, 300, 600, 1800, 3600 seconds for auto logging.
2.
Use the up/down arrow buttons to select the sampling rate (press and hold for quick scroll).
3.
Press the ENTER button to confirm the entry and automatically step to the next parameter.
4.
Press and hold the SET button for 2 seconds to exit to the normal operation mode (or wait 7
seconds for the meter to automatically exit).
Enabling/Disabling the Auto Power OFF Feature (PoFF)
1.
From the
PoFF
parameter screen use the arrow buttons to select ON (enable) or OFF
(disable). With the Auto Power OFF feature enabled, the meter will automatically switch
OFF after 5 minutes of inactivity.
2.
Press ENTER to confirm setting and to step to the next parameter.
3.
If desired, press and hold the SET button for 2 seconds to exit to the normal operation
mode (or wait 7 seconds for the meter to automatically exit).
Set the Beeper Sound ON or OFF (bEEP)
1.
From the
bEEP
parameter window use the arrow buttons to select ON (enable) or OFF
(disable).
2.
Press ENTER to confirm setting and to move to the next parameter.
3.
If desired, press and hold the SET button for 2 seconds to exit to the normal operation
mode (or wait 7 seconds for the meter to automatically exit).
Numerical Format (comma or decimal) ‘dEC’
European and USA numerical formats differ. The meter defaults to USA mode where a decimal
point is used to separate units from tenths, i.e.
20.00
; The European format uses a comma, i.e.
20,00
to separate units from tenths. To change this setting:
1.
From the
dEC
parameter screen use the arrow buttons to select USA or Euro (European).
2.
Press ENTER to confirm setting and to move to the next parameter.
3.
If desired, press and hold the SET button for 2 seconds to exit to the normal operation
mode (or wait 7 seconds for the meter to automatically exit).