Alarm
Delay
The
preset
alarm
delay
time
for
the
Single
Event
Alarm
mode
is
always
zero.
The
adjustable
delay
time
range
for
the
Cumulative
Alarm
is
5
~
120
minutes.
Alarm
Modes
Single
Event
Alarm
mode:
The
LED
immediately
starts
flashing
red
every
10
seconds
when
the
measured
value
exceeds
the
alarm
threshold.
Cumulative
Alarm
mode:
The
LED
begins
flashing
red
only
after
the
average
value
exceeds
the
alarm
threshold
and
also
after
any
programmed
Alarm
Delay
time.
See
Alarm
Delay
parameter
above.
Alarm
Disable
(default):
Alarm
function
disabled.
Alarm
Note:
The
LED
will
continue
flashing
red
every
10
seconds
(once
it
is
triggered
by
an
alarm)
even
when
the
measurement
later
returns
to
the
normal
range.
To
stop
the
LED
alarm,
plug
the
logger
into
the
PC
USB
port
to
generate
a
report,
or
power
off
the
logger
(short
press
of
the
power
button
while
it
is
not
connected
to
the
PC).
Alarm
Limits
Select
the
alarm
threshold
ranges
for
air
and
external
probe
temperature
(TH30)
or
air
temperature
and
RH%
(RHT30).
For
example,
if
the
external
probe
temperature
range
is
set
to
2.0~8.0
o
C,
when
the
measurement
falls
below
2
o
C
or
exceeds
8
o
C
the
LED
will
flash
red
every
10
seconds.
Note
:
If
an
alarm
occurs
between
two
logging
points,
the
logged
data
will
not
show
the
alarm
but
the
LED
may
still
flash
red
every
10
seconds,
especially
in
Single
Event
alarm
mode.
If
desired,
use
the
Cumulative
Alarm
mode
to
minimize
this
effect.
Password
A
16
alphanumeric
character
(max.)
password
may
be
used
to
prevent
unauthorized
re
‐
programming.
User
Company
Name
Input
a
company
name
to
be
included
on
the
log
report
(20
characters
max.).
A
Note
On
Time
Zones
The
logger
automatically
synchronizes
the
time
zone
with
the
connected
PC
when
‘SAVE’
is
selected.
Note:
Time
zone
changes
occurring
during
logging
are
not
taken
into
account.
Once
all
the
programming
is
complete,
press
‘SAVE’
to
confirm
the
settings.
Close
the
setup
window,
remove
the
logger
from
the
PC
USB
port,
and
start
datalogging
as
described
in
the
next
section.