RH350/355    V1.7    11/04 

Minimum (MN) Maximum (MX) Average (AVG) Function 

The MIN/MAX/Average mode allows the user to view only the lowest (

MN

), highest (

MX

) or 

average readings. Press the 

[M

in

/AV

RST

]

 button once and 

MN

 appears on the display. The 

display is now showing the lowest humidity and temperature readings in memory. Press the 

[M

in

/AV

RST

]

 button again and 

MX

 appears on the display. The display is now showing the 

maximum humidity and temperature readings in memory. Press the 

[M

in

/AV

RST

]

 button again 

to display the average (

AVG

) readings in memory. To exit the MIN/MAX/AVG mode, press 

the 

[M

in

/AV

RST

]

 button. To clear the current min/max/avg readings in memory, press and hold 

the

 [M

in

/AV

RST

button for more than two seconds. 

Backlight Display 

Press the 

[M

in

/AV

RST

]

 and 

[MEM]

 buttons together to turn on the backlight. It will remain on 

for ten seconds. 

 

Automatic Power OFF 

With the meter off, press and hold down the 

[ON/OFF

SET

]

 

+

 

[REC

HOLD

]

 buttons for more than 

1 second to enter auto power off settings. Press the 

[MEM]

 button to select non-sleep (n) or 

2, 5, 10, 20, 40 and 60 minutes autosleep. Press 

[T2

WB

]

 to confirm the selection. 

 

Manual 

 

Datalogging

 

Manual datalogging allows the user to store and recall up to 99 readings with the press of a 
button. 

1. Press 

the 

[MEM]

 button to manually store a single reading into memory. The display 

will flash 3 times and the memory location number will be displayed below the humidity 
reading. 

2.  To view stored readings, press and hold the 

[REC

HOLD

]

 button for 2 seconds and 

“RECALL” will blink next to the memory location number on the display. 

3. Press 

the 

[MEM]

 button to manually increment through the stored readings. 

4. Press 

[REC

HOLD

]

 

+

 

[MEM]

 for more than 1 second to clear the memory. 

5. Press 

[REC

HOLD

]

 for more than 1 second to return to normal operation. 

 
Note:

 In RECALL mode, pressing the 

[M

in

/AV

RST

]

 button for min/max/avg readings will auto 

search the stored readings in memory and display the minimum, maximum and 
average values of the stored readings. 

Note

: An infrared printer port is located on the side of the meter. This feature is used on more 

advanced models and is not functional on this model. 

 

Summary of Contents for RH355

Page 1: ...perature T1 T2 Air T1 T1 DP Humidity Air Temperature Dew Point and Wet Bulb Advanced features include built in memory and recall of up to 99 data points Min MAX Average readings from the captured data...

Page 2: ...lculated from RH and Air temperature measurements Low Battery Indicator Yes Overrange Indicator E2 air E5 T1 E8 T2 Underrange Indicator E3 air E6 T1 E9 T2 Operating Conditions 4 to 122 o F 20 to 50 o...

Page 3: ...ween Air temperature T1 temperature Air T1 and DP Dew Point 7 Pressing the T2 WB button toggles the lower left display between WB Wet Bulb T1 T2 T2 and T1 DP Dew Point Selecting temperature units of m...

Page 4: ...enter auto power off settings Press the MEM button to select non sleep n or 2 5 10 20 40 and 60 minutes autosleep Press T2 WB to confirm the selection Manual Datalogging Manual datalogging allows the...

Page 5: ...reading may vary depending on operating temperature will flash on the LCD screen After 30 minutes the flashing will stop to indicate that the 33 calibration is complete Note To exit from the calibrati...

Page 6: ...x xC Hxx x Txxx xC Txxx xC Txxxx xC Txxxx xCLRCCRLF or Txxx xF Hxx x Txxx xF Txxx xF Txxxx xF Txxxx xFLRCCRLF Where The 1 st value is Air Temperature the 2 nd value is Humidity the 3 rd value is Dew P...

Page 7: ...ee of defects in parts and workmanship for one year from date of shipment a six month limited warranty applies on sensors and cables If it should become necessary to return the instrument for service...

Reviews: