OFFSET ADJUSTMENT 

The offset adjustment can be used to correct any known linearity errors with the thermocouple. 

1. 

Select the measure mode and a thermocouple type (J/K/T/E/C/R/S/N). 

2. 

Press and Hold the SET button for 1 second to enter the offset adjust mode. 

3. 

Press the 

 or 

 button to change the offset value. 

4. 

Press and Hold the SET button for 1 second to zero the offset value. 

5. 

Momentarily press the SET button to save the change and exit the function. 

 

 

 

 and 

 

BUTTONS 

The arrow buttons are used to adjust the output value in the SOURCE mode. 

1. 

Select the SOURCE mode 

2. 

Press the 

 or 

 button select a digit for adjustment. The blinking underline curser 

identifies the digit selected. 

3. 

Press the 

or 

button to adjust the value of the digit. Press and Hold the 

or 

 button to 

rapidly adjust the value. 

 

SET BUTTON

 

The SET button is used to manually step through the 5 stored output values. 

1. 

Select the SOURCE mode 

2. 

Press the SET button and the value stored in memory location 01 will be sourced. “MEM.01” 
appears in the display. 

3. 

Each press of the SET button will step through the 5 memory locations.  

4. 

The arrow buttons can be used to adjust the value in each memory location. 

 

STEP/MEM BUTTON

 

The STEP/MEM button is used to automatically step through the 5 stored output values. The meter 
can be set for a single cycle of the stored values or a continuous cycle. 

1. 

Select the SOURCE mode 

2. 

Press and Hold the STEP/MEM button. ”STEPSS” (single cycle) and “STEPSC” (continuous 
cycle) will alternately appear in the display. Release the button when the desired mode is 
displayed. 

3. 

In single cycle mode the meter will source the value displayed in MEM01 for 5 seconds. The 
meter will then advance to MEM02 for 5 seconds. This will continue through MEM05 and 
then step down through the memory locations. The cycle will end when MEM01 has been 
reached. 

4. 

In the continuous mode the cycle will continue until manually stopped. 

5. 

Momentarily press the MEM button to stop the cycle. “END” will briefly appear in the display. 

 

Summary of Contents for PRC30

Page 1: ...USER GUIDE Multifunction Process Calibrator Model PRC30 ...

Page 2: ...ent Double insulation Safety Notes Do not exceed the maximum allowable input range Turn the unit OFF when the device is not in use Remove the batteries if the device is to be stored for longer than 60 days Never dispose of batteries in a fire Batteries may explode or leak Never mix battery types Always install new batteries of the same type Cautions Improper use of this meter can cause damage shoc...

Page 3: ...cklight ZERO button 10 Input Output jacks and Thermocouple mini connector Display Layout 1 SOURCE mode icon 2 ZERO function status icon 3 Thermocouple type 4 MEASURE mode icon 5 Loop Power icon 6 Auto Power OFF active icon 7 Battery status icon 8 Measure mode value 9 Measure mode units icon 10 Datalogger memory location 11 Source mode value 12 Source mode units icon 13 Cold Junction Compensation C...

Page 4: ...ct mA or in the current function or to select mV V in the voltage function voltage is autoranging in the MEASURE mode TYPE BUTTON Press and Hold for 1 second the TYPE UNIT button to change the thermocouple type J K T E C R S N or mV in the temperature function I O BUTTON Momentarily press the I O button to select either SOURCE output or MEASURE input MODE BUTTON Press and Hold the MODE I O button ...

Page 5: ...elect the SOURCE mode 2 Press the SET button and the value stored in memory location 01 will be sourced MEM 01 appears in the display 3 Each press of the SET button will step through the 5 memory locations 4 The arrow buttons can be used to adjust the value in each memory location STEP MEM BUTTON The STEP MEM button is used to automatically step through the 5 stored output values The meter can be ...

Page 6: ... 00mV M4 750 0 o C 1382 0o F 300 0 o C 572 0o F 500 0 o C 932 0o F 40 00mV M5 1000 0o C 1832 0 o F 400 0o C 752 0 o F 750 0o C 1382 0 o F 50 00mV Memory Location mA mV V M1 4 00mA 0 0 0mV 0V M2 8 00mA 25 500mV 5V M3 12 00mA 50 1000mV 10V M4 16 00mA 75 1500mV 15V M5 20 00mA 100 2000mV 20V To change the values in memory 1 Select the SOURCE mode 2 Press the SET button to select the memory location to...

Page 7: ...emperature and the thermocouple type selected The values can be output either manually or stepped from memory as described earlier 1 Turn the meter ON 2 MEASURE will appear in the display 3 Press and Hold the MODE button to select the temperature function 4 Press and Hold the TYPE button to select the thermocouple type 5 Momentarily press the UNIT button to select F or C 6 Momentarily press the I ...

Page 8: ...e current or voltage can be output either manually or stepped from memory as described earlier 1 Turn the meter ON 2 MEASURE will appear in the display 3 Press and Hold the MODE button for 1 second to select a current or voltage function 4 Momentarily press the I O button to select SOURCE 5 Momentarily press the UNIT button to select mA or mV V 6 Connect the Calibration Cable to the meter 7 Connec...

Page 9: ...e battery cover 2 Remove and replace the batteries observing polarity 3 Replace and secure the battery cover All EU users are legally bound by the Battery Ordinance to return all used batteries to community collection points or wherever batteries accumulators are sold Disposal in household trash or refuse is prohibited Disposal Follow the valid legal stipulations in respect of the disposal of the ...

Page 10: ...imum Meter Power 6 AA batteries or AC adaptor Auto Power OFF Meter automatically powers off after 10 minutes of inactivity Operating Temperature 5ºC to 40ºC 41ºF to 104ºF Storage Temperature 20 ºC to 60 ºC 4 ºF to 140 ºF Operating Humidity Max 80 up to 31ºC 87ºF decreasing linearly to 50 at 40ºC 104ºF Storage Humidity 80 Operating Altitude 7000ft 2000meters maximum Dimensions 159 x 80 x 44 mm 6 3 ...

Page 11: ... 3182 F 0 to 1750 C Type N 58 to 2372 F 50 to 1300 C mV Measure 10mV to 60mV Source 5mV to 55mV 0 01mV 0 01 1digit Mode Function Range Resolution Accuracy of reading DC Measure Current 0 to 50mA 0 01mA 0 01 1 digit Percent 25 to 230 0 1 Voltage autoranging 0 to 2000mV 1mV 2 to 20V 0 01V DC Source Current 0 to 24mA 0 01mA Percent 25 to 125 0 1 Voltage 0 to 1999mV 1mV 0 to 20V 0 01V Loop Power Curre...

Reviews: