11
PRC30-en-EU_V1.3 7/14
Range Specifications
T/C Source and Measure Ranges
Resolution
Accuracy
(%
of reading)
Type J
-58 to 1832
°
F (-50 to 1000
°
C)
0.1
°
(measure)
1
°
(source)
± (0.05% + 1
°
C /1.8
°
F)
Type K
-58 to 2498
°
F (-50 to 1370
°
C)
Type T
-184 to 752
°
F (-120 to 400
°
C)
Type E
-58 to 1382
°
F (-50 to 750
°
C)
Type C
32 to 3182
°
F (0 to 1750
°
C)
Type R
32 to 3182
°
F (0 to 1750
°
C)
Type S
32 to 3182
°
F (0 to 1750
°
C)
Type N
-58 to 2372
°
F (-50 to 1300
°
C)
mV
Measure: -10mV to 60mV
Source: -5mV to 55mV
0.01mV
± (0.01% + 1digit)
Mode
Function
Range
(Resolution)
Accuracy
(% of reading)
DC Measure
Current
0 to 50mA (0.01mA)
± (0.01% + 1 digit)
Percent (%)
-25% to +230% (0.1%)
Voltage
(autoranging)
0 to 2000mV (1mV)
2 to 20V (0.01V)
DC Source
Current
0 to 24mA (0.01mA)
Percent (%)
-25% to +125% (0.1%)
Voltage
0 to 1999mV (1mV)
0 to 20V (0.01V)
Loop Power
Current
24 to 30VDC, <50mA
Copyright © 2014 FLIRSystems, Inc.