pH Measurement, Calibration, and Electrode Considerations 

 

Error messages 

ERR-1: Electrode zero potential error

 

and

 

ERR-2: Electrode slope 

error; For either error, check the following:

 

1.  Air bubbles in the electrode bulb. Shake rigorously to remove air bubbles 
2.  Accuracy of the pH buffers used in calibration. Replace buffers if necessary 
3.  Set meter to its factory default state in Parameter P7 (previous section of manual) 

  Calibration intervals depend on the sample, the electrode performance, and the 

required accuracy. For high accuracy measurements (

 ±0.02pH), the meter should be 

calibrated immediately before taking a measurement. For general accuracy (

±0.1pH), 

the meter can be calibrated and used for approximately one week before the next 
calibration. 

  The meter must be recalibrated in the following situations: 

1.  New probe, or probe that is unused for a long period of time 

2.  After measuring acids (pH<2) or alkaline solutions (pH>12) 

3.  After measuring a solution that contains fluoride or a concentrated organic solution 

4.  If the solution’s temperature differs widely from the calibration solution temperature 

  The soaking solution contained in the supplied protective bottle is used to maintain 

activation in the glass bulb and junction. Loosen the capsule, remove the electrode and 
rinse in purified water before taking a measurement. Insert the electrode and tighten 
the capsule after measurements to prevent the solution from leaking. If the soak 
solution is turbid or moldy, replace the solution. 

  To prepare a soak solution: Use 25g pure KCL dissolved with purified water and diluted 

to 100mL. The electrode should not be soaked in a purified water protein solution or an 
acid fluoride solution for long periods of time. In addition, do not soak the electrode in 
organic silicon lipids. 

  For calibration accuracy, the pH of the standard buffer solution must be reliable. The 

buffer solution should be refreshed often, especially after heavy use. 

  For best accuracy, always keep the meter clean and dry, especially the meter’s 

electrode and electrode jack. Clean with medical cotton and alcohol if necessary. 

  The sensitive glass bulb at the front of the combination electrode should not come in 

contact with hard surfaces. Scratches or cracks on the electrode will cause inaccurate 
readings. Before and after each measurement, the electrode should be washed with 
purified water and dried. Do not clean the glass bulb with a tissue for it will affect the 
stability of the electrode potential and increase the response time. The electrode 
should be thoroughly cleaned if a sample sticks to the electrode. Use a solvent if the 
solution does not appear clean after washing. 

  Electrodes that have been used over a long period of time, used in a strong solution 

that has damaged the sensitive bulb, or used with a substance resulting in a jam at the 
junction will be become passivated; the sensitivity will decrease, its response will slow, 
and the readings will be inaccurate. Replace the electrode as soon as possible in these 
cases. 

  For abnormal readings, try calibrating again; if the problem persists replace the 

Electrode. The user can also try resetting the meter to factory default conditions per 
Parameter P7 (detailed in an earlier section). Electrode life can be shortened by heavy 
use, extreme conditions, and improper maintenance. 

www.burntec.com

Summary of Contents for pH300

Page 1: ...User s Guide Digital pH Meter Model pH300 www burntec com ...

Page 2: ...olutions The meter s digital filter improves measurement speed and accuracy Two special modes ammonia and purified water are provided The meter offers a backlit LCD display and is dust proof and water proof meeting the IP57 rating This meter is shipped fully tested and calibrated and with proper use will provide years of reliable service Supplied Equipment List pH300 Meter pH ATC Automatic Tempera...

Page 3: ...4 Electrode connection jack 5 Electrode plug 6 Electrode measuring bulb 7 Electrode body 8 Electrode cable Display Description 1 Parameter Mode icon 2 Measurement reading 3 888 Stored data serial number M measurement to be stored icon RM reading to be recalled icon 4 Unit of measure 5 Temperature reading 6 Stable measurement icon 7 Calibration icons 8 Low battery icon www burntec com ...

Page 4: ...o change the displayed number or the ON OFF state Store Recall and Clear Datalogger Memory STORE Readings The meter can store up to 100 pH and 100 mV readings for a total of 200 data points To store a reading wait until the reading stabilizes the smiling face icon appears when the reading stabilizes Press the M RM key momentarily to store a reading The M icon appears and the data point serial numb...

Page 5: ...on Stir the solution briefly and allow it to stay in the buffer solution until a stable reading is reached 6 Press CAL again and the display will show a blinking 4 00 7 This portion of the calibration procedure is complete when the display stops blinking and shows the C3 icon The unit will automatically switch to the third point of the calibration 8 Rinse the probe in distilled water again allow i...

Page 6: ...series selection USA Europe U S A NIS NIST CH China P2 Distilled water pH temperature compensation setting see note below this table OFF ON P3 Distilled water with Ammonia pH temperature compensation setting see note below this table OFF ON P4 Temperature unit setting P5 Back light display time setting 0 1 3 6min P6 Auto power off setting 0 10 20min P7 Restore to default factory settings OFF ON No...

Page 7: ...parameter P4 or press ENTER to return to the normal measurement mode 3 See the Note below the Table in the previous section for more details on parameter P3 Parameter P4 Temperature Measurement Units Setting 1 From the P4 menu use the CAL or the M RM keys to turn this feature ON or OFF 2 Momentarily press MODE to move to the next parameter P5 or press ENTER to return to the normal measurement mode...

Page 8: ...urified water and diluted to 100mL The electrode should not be soaked in a purified water protein solution or an acid fluoride solution for long periods of time In addition do not soak the electrode in organic silicon lipids For calibration accuracy the pH of the standard buffer solution must be reliable The buffer solution should be refreshed often especially after heavy use For best accuracy alw...

Page 9: ...ompt Parameters Code Settings P1 Back light display time 0 1 3 6 min P2 Auto power off time 0 10 20 min Parameter P1 Display Backlight Setting 1 Press MODE to access the P1 Parameter 2 Use the CAL or the M RM keys to select 0 1 3 or 6 minute default backlighting time 3 Momentarily press MODE to move to the next parameter P2 or press ENTER to return to the normal measurement mode Parameter P2 Autom...

Page 10: ...1999 mV to 0 to 1999mV Resolution 1mV Accuracy Meter 0 1 FS Other Technical Parameters Data storage 200 data groups Storage content Data serial number measurement value unit of measure Power Two AA batteries 1 5V x 2 Size and weight Meter 2 6 x 4 7 x 1 2 65 120 31mm 6 3 oz 180g Case 10 x 8 3 x 2 255 x 210 x 50mm 1 7 lbs 790g Quality Safety certification ISO9001 CE and CMC Working Conditions Enviro...

Page 11: ... is responsible for shipping charges freight insurance and proper packaging to prevent damage in transit This warranty does not apply to defects resulting from action of the user such as misuse improper wiring operation outside of specification improper maintenance or repair or unauthorized modification Extech specifically disclaims any implied warranties or merchantability or fitness for a specif...

Page 12: ...pH300 V1 0 7 09 12 www burntec com ...

Reviews: