background image

pH300   V1.0    7/09 

10

Specifications 

 
pH Specifications 

 

Measuring range 

-2.00 to 19.99 pH 

Resolution 0.1/0.01 

pH 

Accuracy 

Meter: ±0.01pH; Complete Kit: ±0.02pH 

Input current 

2×10

-12 

Input impedance 

1×10

12 

Ω

 

Stability ±0.01 

pH/3h 

Temp. Compensation range 

32 to 212 

o

F (0 to 100 

) (automatic) 

 
mV Specifications 

 

Measuring range (mV/E

H

-1999 mV to 0 to 1999mV 

Resolution 1mV 

Accuracy 

Meter: ±0.1% FS 

 

Other Technical Parameters 

 

Data storage 

200 data groups 

Storage content 

Data serial number, measurement value, unit of measure 

Power 

Two AA batteries (1.5V x 2) 

Size and weight 

Meter: 2.6 x 4.7 x 1.2” (65 × 120 × 31mm) / 6.3 oz (180g)  

Case: 10 x 8.3 x 2” (255 x 210 x 50mm) / 1.7 lbs (790g) 

Quality / Safety certification 

ISO9001, CE and CMC 

 
Working Conditions 

 

Environment temperature 

41 to 95 

o

F (5 to 35 

Environmental humidity 

85% 

IP rating 

IP57 Dustproof and waterproof 

 

 

www.burntec.com

Summary of Contents for pH300

Page 1: ...User s Guide Digital pH Meter Model pH300 www burntec com ...

Page 2: ...olutions The meter s digital filter improves measurement speed and accuracy Two special modes ammonia and purified water are provided The meter offers a backlit LCD display and is dust proof and water proof meeting the IP57 rating This meter is shipped fully tested and calibrated and with proper use will provide years of reliable service Supplied Equipment List pH300 Meter pH ATC Automatic Tempera...

Page 3: ...4 Electrode connection jack 5 Electrode plug 6 Electrode measuring bulb 7 Electrode body 8 Electrode cable Display Description 1 Parameter Mode icon 2 Measurement reading 3 888 Stored data serial number M measurement to be stored icon RM reading to be recalled icon 4 Unit of measure 5 Temperature reading 6 Stable measurement icon 7 Calibration icons 8 Low battery icon www burntec com ...

Page 4: ...o change the displayed number or the ON OFF state Store Recall and Clear Datalogger Memory STORE Readings The meter can store up to 100 pH and 100 mV readings for a total of 200 data points To store a reading wait until the reading stabilizes the smiling face icon appears when the reading stabilizes Press the M RM key momentarily to store a reading The M icon appears and the data point serial numb...

Page 5: ...on Stir the solution briefly and allow it to stay in the buffer solution until a stable reading is reached 6 Press CAL again and the display will show a blinking 4 00 7 This portion of the calibration procedure is complete when the display stops blinking and shows the C3 icon The unit will automatically switch to the third point of the calibration 8 Rinse the probe in distilled water again allow i...

Page 6: ...series selection USA Europe U S A NIS NIST CH China P2 Distilled water pH temperature compensation setting see note below this table OFF ON P3 Distilled water with Ammonia pH temperature compensation setting see note below this table OFF ON P4 Temperature unit setting P5 Back light display time setting 0 1 3 6min P6 Auto power off setting 0 10 20min P7 Restore to default factory settings OFF ON No...

Page 7: ...parameter P4 or press ENTER to return to the normal measurement mode 3 See the Note below the Table in the previous section for more details on parameter P3 Parameter P4 Temperature Measurement Units Setting 1 From the P4 menu use the CAL or the M RM keys to turn this feature ON or OFF 2 Momentarily press MODE to move to the next parameter P5 or press ENTER to return to the normal measurement mode...

Page 8: ...urified water and diluted to 100mL The electrode should not be soaked in a purified water protein solution or an acid fluoride solution for long periods of time In addition do not soak the electrode in organic silicon lipids For calibration accuracy the pH of the standard buffer solution must be reliable The buffer solution should be refreshed often especially after heavy use For best accuracy alw...

Page 9: ...ompt Parameters Code Settings P1 Back light display time 0 1 3 6 min P2 Auto power off time 0 10 20 min Parameter P1 Display Backlight Setting 1 Press MODE to access the P1 Parameter 2 Use the CAL or the M RM keys to select 0 1 3 or 6 minute default backlighting time 3 Momentarily press MODE to move to the next parameter P2 or press ENTER to return to the normal measurement mode Parameter P2 Autom...

Page 10: ...1999 mV to 0 to 1999mV Resolution 1mV Accuracy Meter 0 1 FS Other Technical Parameters Data storage 200 data groups Storage content Data serial number measurement value unit of measure Power Two AA batteries 1 5V x 2 Size and weight Meter 2 6 x 4 7 x 1 2 65 120 31mm 6 3 oz 180g Case 10 x 8 3 x 2 255 x 210 x 50mm 1 7 lbs 790g Quality Safety certification ISO9001 CE and CMC Working Conditions Enviro...

Page 11: ... is responsible for shipping charges freight insurance and proper packaging to prevent damage in transit This warranty does not apply to defects resulting from action of the user such as misuse improper wiring operation outside of specification improper maintenance or repair or unauthorized modification Extech specifically disclaims any implied warranties or merchantability or fitness for a specif...

Page 12: ...pH300 V1 0 7 09 12 www burntec com ...

Reviews: