background image

8                                             MC108  V1.1   2/09 

 

Click 

 and 

 icon to permanently erase the image files inside the Flash-Memory. 

 

 

 

Figure 2. 

 
The right lower portions of the screen both have the same function in the VT101 

Click the

 icon for real-time imaging.The observed image is shown as figure 3. You 

may click the “Show Magnification” or “Show Ruler” option, the effect is also shown as 
figure 3. Please remember ruler and magnification number are for reference only. They are 
shown when the optical zoom/focus indicator is located at text 7 and 27. 
Magnification is variable by clicking the

  or 

icon. Transmission may be slow, after 

clicking the icon; you may have to wait a moment to obtain the desired image. 

Click the

 icon in figure 3 to save the display image to PC. 

Click the

 icon to leave Continuous Image Transfer mode. 

 

 

Figure 3. 

Click the 

 icon will disconnect the microscope from the PC. 

Click the 

 icon will leave the program. 

 
 
 

Summary of Contents for MC108

Page 1: ...User s Guide Mini Microscope Model MC108 ...

Page 2: ...e in areas with high humidity Use the microscope only within its operating temperature range 23 F to 122 F 5 C to 50 C Sudden varied temperatures changes may cloud the lens Do not let the lens face directly into sunlight or intense light for an extended period of time Strong light can damage the light sensitive electronic circuits Avoid touching the lens Caution The white LEDs have strong luminosi...

Page 3: ...to maintain battery strength this is suggested to extend battery longevity For lengthy storage periods remove batteries to prevent damage due to battery leakage discharge The unit may also use alkaline batteries Do not mix battery types in use and do not recharge alkaline batteries Powering the Unit Press the power button for 2 seconds and then enter the normal observation Mode The magnified image...

Page 4: ...zoom wheel to save the image into RAM The unit can enlarge image to 108x 27x4 108x Aiming Mark Circular mark on the back of case shows position of lens which helps locate lens over the target object Special Effects Press this button to select different image effects E g normal gray inverse ruler and comparison Power Hold this button for 2 seconds to turn the unit on off Tools Press this button to ...

Page 5: ...l settings Image Effects Press the Image Effects button to to change the appearance of the image The image mode selected is indicated by the corresponding letter next to the battery symbol in the display N G I E or D Normal N Normal display Gray G To remove color and create a gray scale image for easier image discrimination Inverse I To reverse color and or grayscale to help highlight some given f...

Page 6: ...o leave this mode Status Battery Low Battery power is low Image Effect N Normal Mode G Gray Mode I Inverse Mode E Emboss Mode D Dual Window View Mode Freeze When in Dual Window View mode or Dual Window Compare mode Pressing Confirm button will enter Freeze mode Press Confirm button again to leave Freeze mode Magnifying Current magnification rate is shown when the indicator is located at text 7 and...

Page 7: ...mation please refer to the website www silabs com CP2102 is a product of SILICON LABS Inc You may use the add remove programs on the control panel to remove the application program and USB driver Application Program Instructions When starting the application program see figure 1 click the Language to select your preference then click the icon A Digital Microscope dialog will appear on the screen c...

Page 8: ...wn as figure 3 Please remember ruler and magnification number are for reference only They are shown when the optical zoom focus indicator is located at text 7 and 27 Magnification is variable by clicking the or icon Transmission may be slow after clicking the icon you may have to wait a moment to obtain the desired image Click the icon in figure 3 to save the display image to PC Click the icon to ...

Page 9: ...FLIR COMPANY warrants this instrument to be free of defects in parts and workmanship for one year from date of shipment a six month limited warranty applies to sensors and cables If it should become necessary to return the instrument for service during or beyond the warranty period contact the Customer Service Department at 781 890 7440 ext 210 for authorization or visit our website www extech com...

Reviews: