CG104-en-GB_V1.8 1/17
9
Specifications
Electrical Specifications
Ferrous
Non-Ferrous
Measurement
principle
Magnetic induction
Eddy current principle
Response time
One second
One second
Detectable Substrate
Materials
Iron, Steel
Copper, Aluminum, Zinc,
Bronze, Brass, and others
Measuring range
0~2000
μ
m
0~80.0mils
0~1020
μ
m
0~40.0mils
Accuracy
1
(% of reading)
0~199
μ
m: ±(10 dgts)
200~1000
μ
m: (±3.0% + 10 dgts)
1001~1999
μ
m: (±5.0% + 10 dgts)
0~7.8mils: ±(4dgts)
7.9~39.0mils: ±(3% + 4 dgts)
39.1~80.0mils: (±5% + 4 dgts)
0~199
μ
m: ±(10 dgts)
200 ~1020
μ
m: (±3.0% + 10dgts)
0~7.8mils: ±(4 dgts)
7.9~40.0mils: (±3.0% + 4 dgts)
Resolution
0~100
μ
m: (1
μ
m)
100
μ
m~1000
μ
m: (1
μ
m)
1000
μ
m~2000
μ
m: (1
μ
m)
0~10mils: (0.1mils)
10mils~80mils: (0.1mils)
0~100
μ
m: (1
μ
m)
100
μ
m ~1020
μ
m: (1
μ
m)
0~10mils: (0.1mils)
10mils~40.0mils: (0.1mils)
Basic critical thickness
0 to 80mils (0 to 2000um)
0 to 40mils (0 to 1020um)
Industrial standards
Conforms to GB/T 4956-1985, GB/T 4957-1985, JB/T 8393-1996, JJG
889-95, and JJG 818-93
Temperature
Coefficient
0.1 times the applicable accuracy specification per °C from 18°C to
28°C (64°F to 82°F)
1
Accuracy note: Accuracy statement applies to use on a flat surface, with a zero and a
calibration performed near the thickness of the film to be measured, with an identical base
metal and with the meter stabilized at ambient temperature 18 to 28°C (64 to 82°F).
The accuracy of the reference films or any reference standards should be added to
measurement results.
www.
GlobalTestSupply
.com
Find Quality Products Online at: