Maintenance 

CLEANING AND STORAGE 

1.  The meter should be cleaned with a damp cloth and mild detergent when necessary. Do not use 

solvents or abrasives. 

2.  Store the meter in an area with moderate temperature and humidity. 

 

 

Specifications 

 

Function Range 

Resolution 

Accuracy 

CO

2

 

0 to 9999ppm 

1ppm 

±(5%rdg +50ppm) 

Temperature 

-10 to 60°C 
14 to 140°F 

0.1° ±0.6°C/0.9°F 

Humidity 

0.1 to 99.9% 

0.1% 

±3%(10 to 90%) 
±5%(< 10% or > 90%) 

 

Display LCD 

 

Sensor Type 

CO

2

: NDIR (non-dispersive infrared) technology  

Humidity: Capacitance sensor;  
Temperature (air): Thermistor 

Response CO

2

: <2min for 90% step change 

Temp: <2min for 90% step change 
%RH: <10min for 90% step change 

Datalogging 

Up to 15999 points 

Sample rate 

1 second to 4 hours-59 minutes-59 seconds 

Operating Conditions 

-10 to 60°C (14 to 140°F); < 90% RH non-condensing 

Storage Conditions 

-20 to 60°C (-4 to 140°F); <99% RH non-condensing 

Power Supply 

5VDC (±10%), 

 500mA (AC adaptor supplied) 

Dimensions / Weight 

117x102x102mm (4.6x4x4”); 204g (7.2 oz.) 

 
 
 

Summary of Contents for 793950502105

Page 1: ...User s Guide CO2 Monitor and Datalogger Model CO210...

Page 2: ...Date and Time 3 Air Temperature 4 Relative Humidity Symbols ppm CO2 value GOOD CO2 air quality level NORMAL CO2 l air quality level POOR CO2 air quality level Air Temp Air Temperature Humidity Relati...

Page 3: ...he RESET button for more than 1 second to clear the minimum and maximum value from memory DATA LOGGING The meter can record readings of CO2 temperature and humidity for long time environment monitorin...

Page 4: ...ode for more than 1 second to enter the setup mode P1 1 CO2 ALARM GOOD LEVEL When entering the setup mode P1 0 and CO2 are displayed on the LCD Press the SET button again to go into P1 1 for setting t...

Page 5: ...ange the units Press the SET button to confirm the setting or press the ESC button to exit without saving and return to P2 0 P3 0 REAL TIME CLOCK Press the LOG button in P1 0 twice to access P3 0 for...

Page 6: ...settings Parameter Default P1 1 700ppm P1 2 1000ppm P1 3 1000ppm P2 1 C P4 1 No P5 0 SAMPLING RATE Press the LOG button in P1 0 three times to access P5 0 to set the logging sample rate The available...

Page 7: ...ation is suggested to be done outdoors on a sunny day with good ventilation and fresh air where the CO2 level is approximately 400 ppm Do not calibrate on a rainy day because high humidity will affect...

Page 8: ...Humidity 0 1 to 99 9 0 1 3 10 to 90 5 10 or 90 Display LCD Sensor Type CO2 NDIR non dispersive infrared technology Humidity Capacitance sensor Temperature air Thermistor Response CO2 2min for 90 step...

Page 9: ...if it still appear re calibrate the meter If above two methods faile return for repair E17 ABC mode of CO2 senor has failed and might cause wrong CO2 readings Send back for repair Temp Display E02 Air...

Page 10: ...lead to serious oxygen deprivation resulting in permanent brain damage coma and even death Regulatory exposure limits ASHRAE Standard 62 1989 1000ppm CO2 concentration in occupied building should not...

Reviews: