5.6 Frequency Measurements  

 

Note:

 Clamp (ammeter) and test lead (voltage) frequency measurements can both be 

made. Check the specifications for differences in sensitivity. When reading frequency via 

the clamp jaw the current must be at least 5A (minimum threshold). 

 

1. Set the Rotary Selector switch to the Hz function. 

2. Connect the black test lead to the COM terminal and red test lead to the V/

/uF terminal 

and connect the other ends of the test leads to the circuit, component, or other device 

under test for test lead measurements; otherwise use the clamp jaw for current 

measurements. 

3. Read the displayed measurement value on the LCD. 

 

 

5.7 Resistance Measurements  

 

Warning:

 Before attempting any in-circuit resistance measurements remove power from 

the circuit under test and discharge any capacitors. 

 

1. Set the Rotary Selector switch to the 

 function. 

2. Connect the black test lead to the COM terminal and red test lead to the V/

/uF terminal 

and connect the other ends of the test leads to the circuit, component, or other device 

under test. 

3. Read the displayed measurement value on the LCD. 

 

5.8 Capacitance Measurements 

 

Warning:

 Discharge capacitors and circuit before testing. 

 

1. Set the Rotary Selector switch to the uF function. 

2. Remove at least one leg of a capacitor from its circuit before testing. 

3. Connect the black test lead to the COM terminal and red test lead to the V/

/uF terminal 

and connect the other ends of the test leads to the circuit, component, or other device 

under test. 

4. If measuring a polarity sensitive electrolytic capacitor, touch red test lead to the positive 

side of the cap. and the black test lead to the negative side. 

5. Read the displayed measurement value on the LCD. 

 

5.9 Temperature Measurements  

 

1. Set the Rotary Selector switch to the “C / F” function. 

2. Set the temperature units with the temperature units key. 

3. Insert the type-K thermocouple adapter into the meter (positive end to the V/

/uF meter 

terminal). 

4. Insert the thermocouple into the thermocouple adapter. 

5. Secure the beaded end of the K-type thermocouple to the surface to be measured. 

6. Read the displayed temperature value on the LCD. 

 

5.10 Continuity Measurements  

 

Warning:

 Before attempting any in-circuit measurements remove power from the circuit 

under test and discharge any capacitors. 

 

1. Set the Rotary Selector switch to the diode/continuity function (diode symbol). 

2. Connect the black test lead to the COM terminal and red test lead to the V/

/uF 

terminal and connect the other ends of the test leads to the circuit, component, or other 

device under  test. 

3. When the resistance measured is lower than 40

, the meter will sound an audible tone. 

Summary of Contents for 380973

Page 1: ...nusoidal waveform measurements Peak Hold and Data Hold Selectable temperature measurement units Frequency measurements using the current jaws or using the test leads 1 INTRODUCTION Congratulations on your purchase of Extech s 380973 True RMS AC Clamp on DMM This professional meter with proper care will provide years of safe reliable service ...

Page 2: ...1MΩ DC Voltage 600V 1V 0 75 2d 660V rms input imp 1MΩ Resistance 4KΩ 0 001kΩ 1 0 5d 660V rms 3 2VDC open cir Frequency 4KHz clamp test 1Hz 0 5 rdg 5d 800A sensitiv 5A 4KHz lead test 1Hz 0 5 rdg 5d 660V sensitiv 1V Capacitance 2000uF 1uF 3 5d 250V rms Type K Therm 20 to 300o C 1o C 2 3o C 300 to 1200o C 1o C 3 5 2o C 4 to 2192o F 1o F 3 8o F 660 Vrms Continuity Buzzer 40 ohms oc V 3V sc A 0 8mA 660...

Page 3: ...6 Data Hold key Press to activate press again to deactivate 7 Rotary selector switch Powers meter while selecting function and range 8 Negative input terminal Black test lead connection COM 9 Positive input terminal Red test lead connection for voltage resistance frequency capacitance temperature diode and continuity tests 10 Temperature units select key Press to select desired units 4 INTERNATION...

Page 4: ...rmer jaw around one of the conductors under test Open the jaw via the jaw trigger 3 Ensure that the clamp closes completely around the conductor 3 Read the displayed measurement value on the display 5 4 DC Voltage Measurements Warning Maximum DC input voltage is 600V To avoid electrical shock or damage to the instrument do not attempt to measure any voltage exceeding this limit 1 Set the Rotary Se...

Page 5: ... to the uF function 2 Remove at least one leg of a capacitor from its circuit before testing 3 Connect the black test lead to the COM terminal and red test lead to the V Ω uF terminal and connect the other ends of the test leads to the circuit component or other device under test 4 If measuring a polarity sensitive electrolytic capacitor touch red test lead to the positive side of the cap and the ...

Page 6: ...nce 5 The display will now indicate P for Peak Hold or H for Data Hold 6 For Peak Hold the display will show the maximum peak of a measured signal surge For example in rush current during a motor start up can be easily detected using Peak Hold mode For Data Hold the LCD reading is frozen 7 To return to normal operation press the Peak or Data Hold key again 6 MAINTENANCE 6 1 Battery Replacement War...

Page 7: ...440 for authorization A Return Authorization RA number must be issued before any product is returned to Extech The sender is responsible for shipping charges freight insurance and proper packaging to prevent damage in transit This warranty does not apply to defects resulting from action of the user such as misuse improper wiring operation outside of specification improper maintenance or repair or ...

Reviews: