Remote Control Commands
24
T100S-HP Programming Guide
2.5.1.4
F
Syntax
F=xxxxxx.x
Parameter
•
xxxxxx.x
: optical frequency in GHz. The possible wavelength range is available by
using the
command (see
Description
Sets the emission wavelength (in GHz) and automatically switches the displayed unit to
GHz.
The laser is tuned directly to the desired optical frequency without step-mode scanning
or swept-mode hop-free scanning. The tuning is performed at the
MOTOR SPEED
speed
Response
•
RS-232C:
OK
•
GPIB: none, see section
Checking System Status with the Status Byte, p. 9
2.5.1.5
F?
Syntax
F?
Parameter
None.
Description
Returns the value of the present optical frequency value in GHz.
Response
F=xxxxxx.x
2.5.1.6
F? MIN|MAX
Syntax
F? MIN|MAX
Parameter
•
MIN
: minimum operating frequency.
•
MAX
: maximum operating frequency
Description
Returns the minimum or maximum operating frequency of the T100S-HP in GHz.
Response
F=xxxx.xxx
2.5.1.7
MOTOR_SPEED
Syntax
MOTOR_SPEED=xxx
Parameter
xxx
: sweep speed in nm/s, in the range 1 to 100 nm/s. Operational sweep speeds are: 1,
2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, 17,18, 20, 22, 25, 29, 33, 40, 50, 67, 100 nm/s.