Remote Control Commands
30
T100S-HP Programming Guide
2.6
Calibration Control
2.6.1
AUTO_CAL
Syntax
AUTO_CAL
Parameter
None.
Description
Launches the internal wavelength referencing sequence.
Response
•
REFERENCING ERROR
: if auto-calibration has failed, this message is displayed for a
few seconds, and the T100S-HP resumes normal operating status. Unlike the
2nd + I
command (in manual mode) that stalls the system when a referencing error occurs,
there is no need to press
Enter
to restore the system.
•
OK
: (RS-232C only) the calibration has been successfully performed.
2.6.2
PCAL1
Syntax
PCAL1=xx.xx
Parameter
•
xx.xx
: the calibration power at first wavelength, in the range 0.3 to 0.6 mW.
Description
Sets the first power value (in mW) of the two-point power calibration method.
Response
•
RS-232C:
OK
•
GPIB: none, see section
Checking System Status with the Status Byte, p. 9
2.6.3
PCAL2
Syntax
PCAL2=xx.xx
Parameter
•
xx.xx
: the calibration power at second wavelength, in the range 0.3 to 0.6 mW.
Description
Sets the second power value (in mW) of the two-point power calibration method.
Response
•
RS-232C:
OK
•
GPIB: none, see section
Checking System Status with the Status Byte, p. 9
2.6.4
PCAL1?
Syntax
PCAL1?
Parameter
None.