SCPI Command Reference
Multi-Wavelength Meter
137
Command Description
:SENSe[1..n]:AVERage:COUNt
Description
This command sets the number of scans for
averaging.
At *RST, this value is 1.
Syntax
:SENSe[1..n]:AVERage:COUNt<wsp><AverageC
ount>|MAXimum|MINimum
Parameter(s)
AverageCount:
The program data syntax for <AverageCount> is
defined as a <numeric_value> element. The
<AverageCount> special forms MINimum and
MAXimum are accepted on input.
MINimum allows to set the instrument to the
smallest supported value.
MAXimum allows to set the instrument to the
greatest supported value.
The <AverageCount> parameter is the number
of scans used to obtain the average for the final
scan.
Example(s)
/* These lines start a scan in Custom mode */
SENS:AVER:COUN:AUTO OFF
SENS:AVER:COUN 12
INIT:IMM
See Also
SENSe[1..n]:AVERage:COUNt?
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for IQS-5320
Page 9: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 13: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 69: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 91: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 99: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 109: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 113: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 119: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 167: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...