Working with the Tunable Laser Source
22
IQ-2600B
Setting the Sweep Parameters
Figure 3-9.
Synchronization Pattern (Continuous Mode)
(a) sweep start
(b) second sweep trigger
(c) trig received during the second sweep; kept in memory and will trigger
the third sweep
(d) trig received during the third sweep kept in memory and will trigger the
fourth sweep
(e) lost trig
Trigger Option in Stepped Mode
The sweep will begin when you press the
Sweep
button
(F6)
or use the
following remote command: “SOUR:SWE:STAT 1”. The system will position
itself at the start wavelength and wait for a synchronization signal. Then,
the system will wait for a synchronization signal to go to the next step in the
sweep.
Stop
Start
Trig In
t
V
(a)
(b) (c)
(d)(e)
Wavelength
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com