background image

FOT-930 MaxTester

SPECIFICATIONS

1

External Power Meter

FOT-932

FOT-932X

FOT-933

Detector type

Ge

GeX

InGaAs

Measurement range (dBm)

10 to 

70

26 to 

55

6 to 

73

Range displayed (dBm)

Down to 

77

Down to 

65

Down to 

80

Uncertainty

2, 3

± 5 % ± 0.1 nW

± 5 % ± 3 nW

± 5 % ± 0.05 nW

Wavelength range (nm)

800 to 1650

800 to 1650

800 to 1650

Display resolution

2

(dB)

0.01

0.01

0.01

Calibrated wavelengths

40

42

40

Recommended recalibration period (years)

3

3

3

Automatic offset nulling

4

Yes

Yes

Yes

Measurement-distance units                                             kilometers, meters, kilofeet, feet, miles 

Sources

Standard

4

5

12C (second port)

12D (second port)

Wavelengths

5

(nm)

1310 ± 20

1310 ± 20

1310 ± 20

850 ± 25

850 ± 25

1550 ± 20

1550 ± 20

1490 ± 10

1300 +50/–10 

1300 +50/–10 

1625 ± 10

1550 ± 20

Emitter type

Laser 

Laser

Laser

LED

LED

Minimum output power

5

(dBm)

1/

1

1/

4/

7

1/

7/

4

30/

30 (50/125 µm)

24/

24 (62.5/125 µm)

Spectral width

6

(nm)

5/

5/

5/

5

5/

5/

5

50/135

50/135

Stability

7

(8 hours) (dB)

± 0.05

± 0.05

± 0.05

± 0.05

± 0.05

FasTesT

Standard

4

5

12C (second port)

12D (second port)

Wavelengths (nm)

1310 

1310

1310 

850 

850 

1550 1550 

1490

1300 1300 

1625 1550 

Loss range

8

(dB)

60

56

56

40

46

Loss precision

9

(repeatability) (dB)

side-by-side

0.15

0.15

0.15

0.15

0.15

loopback

0.25

0.25

0.25

0.25

0.25

Length measurement range (km)

200

200

200

5

5

Length measurement uncertainty

10

± (10 m + 1 % x length)

Dedicated ORL 

All SM Wavelengths

ORL range (APC / UPC) (dB)

65/55                   

ORL uncertainty

11

(dB)

± 0.5       

Resolution

2

(dB)

0.01

General Specifications

Size (H x W x D) 

25.0 cm x 12.5 cm x 7.5 cm 

(9 7/8 in x 4 15/16 in x 3 in)

Weight 

1 kg

(2.2 lb)

Temperature  operating 

–10 °C to 50 °C 

(14 °F to 122 °F)

storage

13

–40 °C to 70 °C 

(–40 °F to 158 °F)

Storage 

Capacity of 1024 complete tests 

Relative humidity 

0 % to 95 % non-condensing

Power

9

Li-ion battery (9 hours)
3 hours to fully recharge when unit is off

Warranty (years)

3

Standard Accessories

User guide, AC adapter/charger, 2 Li-ION batteries, shoulder strap, Certificate of Calibration.

Talk Set

Emitter type

Laser

Wavelength (nm)

1550 ± 20   

Dynamic range at 1550 nm (dB)

45

Dynamic range MM

12

(dB)

40

VFL

9

Emitter type

Laser

Wavelength (nm)

650

Output power (dBm)

3

Notes

1. At 23 °C ± 1 °C and 1550 nm with FC connector and on batteries, unless otherwise 

specified. 

2. Resolution, uncertainty and linearity are functions of input power; uncertainty is valid at 

calibration conditions.

3. Traceable to NIST; up to 20 dBm for GeX.
4.  Power of > 

45 dBm for Ge, > –30 dBm for GeX and > –57 dBm for InGaAs.

5.  In High source mode.
6.  As defined by Telcordia TR-TSY-000887, rms for lasers and at 

3 dB for LEDs; typical 

values for LEDs.

7. After a warm-up time of 6 minutes, in CW source mode.
8.  Typical value, at 1550 nm for SM and 850 nm for MM.
9. Typical value.
10. For fiber length 

120 km.

11. Typical value.
12. For graded-index MM fibers, typical.
13. Without batteries.

Summary of Contents for FOT-930 MaxTester

Page 1: ...he testing of passive optical networks PONs at 1310 nm 1490 nm and 1550 nm the three wavelengths recommended by the ITU T G 983 3 for PONs Cost of ownership lowest in the industry thanks to three year warranty and recommended calibration interval error free testing and minimized training time Telecommunications Test and Measurement www megahz com MULTIFUNCTION LOSS TESTER FOT 930 MaxTester 930 NET...

Page 2: ...t up to three singlemode wavelengths Bidirectional loss testing at two multimode wavelengths Fiber length measurement Flexible solution five wavelength multimode and singlemode configurations meeting the requirements of installers contractors for all test situations Up to three singlemode wavelengths 1310 1550 and a choice between 1490 and 1625 nm on one port Two multimode wavelengths 850 and 1300...

Page 3: ...ta according to FTTx terminology Similar test data presentation regardless of the location of master and remote units Integrated Data Storage Management This feature enables the FasTesT initiator to save results on a remote unit even when multiple remote units are used Key benefits include The possibility to store test data in a single unit Easier data post processing and transfer from the FTB 393...

Page 4: ...spection probe for easy viewing of connectors and fiber ends on the FOT 930 s high resolution display GeX detector for high power measurement up to 26 dBm Complete report generation capabilities Talk set and VFL options Ease of use for faster testing reduced training minimum error potential etc In 10 seconds the MaxTester s FasTest function provides insertion loss and ORL values for up to three wa...

Page 5: ...er detector talk set visual fault locator VFL and video fiber inspection probe No offset nulling required Internal InGaAs FasTesT detector While performing FasTesT measurements the FOT 930 can launch automated loss and ORL measurements on all three wavelengths and perform fiber length measurements The FOT 930 s Windows like 320 x 240 color interface provides first class user friendliness The FOT 9...

Page 6: ...complete high quality test documentation Its data logging and management features help users quickly access and download test results to any PC through the RS 232 port for in depth analysis and first class report generation Online Help Menu and Multilingual Interface for Enhanced User Friendliness The FOT 930 MaxTester features a comprehensive easy to use on line help menu providing all the necess...

Page 7: ...ainty10 10 m 1 x length Dedicated ORL All SM Wavelengths ORL range APC UPC dB 65 55 ORL uncertainty11 dB 0 5 Resolution2 dB 0 01 General Specifications Size H x W x D 25 0 cm x 12 5 cm x 7 5 cm 9 7 8 in x 4 15 16 in x 3 in Weight 1 kg 2 2 lb Temperature operating 10 C to 50 C 14 F to 122 F storage13 40 C to 70 C 40 F to 158 F Storage Capacity of 1024 complete tests Relative humidity 0 to 95 non co...

Page 8: ... 1550 1625 nm FOT 932X 5 GeX detector triple wavelength 1310 1490 1550 nm FOT 933 InGaAs detector dual wavelength 1310 1550 nm FOT 933 4 InGaAs detector triple wavelength 1310 1550 1625 nm FOT 933 5 InGaAs detector triple wavelength 1310 1490 1550 nm FTB 932 MM Ge detector with multimode second source only 5 If VFL option is available LASER RADIATION AVOID DIRECT EYE EXPOSURE CLASS 3R LASER PRODUC...

Reviews: