Reverting to Factory Settings
To transfer test configurations from one unit or computer to
another:
1.
Using Windows Explorer, on the unit where the custom configurations
are located, select which files you want to transfer. The files are
located in different folders depending on the unit you are using.
In Windows XP on a computer, or on an FTB-500 platform:
C:\Documents and Settings\All Users\Application
Data\EXFO\ConnectorMax\Test configuration files\Custom.
In Windows Vista or 7 on a computer:
C:\ProgramData\Exfo\
ConnectorMax\Test configuration files\Custom.
On an FTB-1 or FTB-200 v2 unit:
D:\Documents and Settings\All
Users\Application Data\EXFO\ConnectorMax\Test configuration
files\Custom.
2.
If the two units are linked through a network and that you have access,
you can paste the files directly into the corresponding folder of the
destination unit. Otherwise, copy the files onto a USB key, then paste
the files in the destination folder once you have.
Reverting to Factory Settings
At any time in the application, you can revert to factory settings in your
menus. However, the
Restore Factory Settings
button is valid only for the
window or tab where you use it.