The following are some common messages that may be displayed, including error messages. If an

error message other than those listed below is displayed, please contact technical support in the USA

at (803) 329-0162 (ext. 0).

eXact

®

 Micro 7+ Meter Messages

LCD Message

HI

LO

LO

ER

Description

In READ mode: test sample concentration is above
WKHPHDVXUHPHQWUDQJHWHVWVSHFL¿F
In READ mode: test sample concentration is below
WKHPHDVXUHPHQWUDQJHWHVWVSHFL¿F
In ZERO mode: sample absorbance (due to a
cloudy or colored sample or a dirty cell) is too high
to zero, the meter will read “LO”.
Excessive stray light detected. Normally this
does not occur, even when testing in sunlight.

Low battery indication.

Corrective Action

Dilute and retest. Dilution Kit available
(Part Number 487200).
Sample value is below measurement 
range.
'LOXWHVDPSOH¿OWHUVDPSOHRUFOHDQFHOO2QHRI
these options should remedy the problem.

Place the LIGHT BLOCKING CAP over the CELL
for zeroing and for reading result. Moving
WRDVKDGHGDUHDFDQDOVR¿[WKLVSUREOHP
Replace the batteries.

About The Built-In Cell

The built-in 

CELL

LVWUDQVSDUHQWSODVWLFDQGZKHQ¿OOHGWRWKHWRSFRQWDLQVPO7KHVWXUG\

CELL 

design 

will last for over 20,000 readings. Scratches on the 

CELL

 will not interfere or compromise the accuracy of 

WKHUHDGLQJVEHFDXVHRILWV¿[HGSRVLWLRQ)RUEHVWDFFXUDF\ULQVHFHOOZLWKFOHDQZDWHULPPHGLDWHO\DIWHU

a test is completed. Do not use solvents, such as acetone, to clean the cell. When the 

CELL

 becomes 

stained or cloudy from repeated testing, or when the meter does not blank when you press the 

ZERO/ON

 

button, the cell needs to be cleaned. Clean as follows)LOOFHOOZLWKFOHDQZDWHUDQGPRYHWKH

Cell 

cleaning brush

 up-and-down and back-and-forth along the walls of the cell. Afterwards, rinse the cell and 

the meter is ready for use again. Cleaning the cell regularly is especially recommended after you run a 

test that is using turbidity or precipitation chemistry for analysis (Calcium Hardness and Cyanuric Acid). 

eXact

®

 Photometer 2-Year Limited Warranty

Registration of your eXact

®

 photometer must be received within 30 days from date of purchase to 

activate the warranty. The eXact

®

 photometer is warranted to be free from defects in materials and 

workmanship for a period of two (2) years from the date of purchase by the customer. ITS will repair or 

replace any part of the product which is deemed to be faulty or otherwise defective. The non-transferable 

warranty does not cover product damage caused by abuse (such as crushing a tablet in the cell) or 

improper use. If the meter is faulty or otherwise defective contact ITS by phone (+1-803-329-9712 Ext. 0) 

or email ([email protected]) to describe the problem and obtain a return authorization form before 

returning the photometer to ITS. Damage caused by improper packing of the photometer for return 

shipment to ITS will not be covered by the warranty. Customer is responsible for shipping charges to 

ITS. ITS pays postage when photometer is returned to customer. A maximum processing fee of $75 will 

be charged for repair or replacement of non-registered photometers and damages not covered by this 

warranty. Registration is available over the phone (+1-803-329-9712 Ext. 0) or online at 

KWWSZZZVHQVDIHFRPPLFURZDUUDQW\3HUVRQDOGDWDLVNHSWFRQ¿GHQWLDO

1. Unscrew the O-ring sealed battery cover counter-clockwise. Use proper sized pliers if necessary.

Do not disturb the sealing O-ring. Batteries are not included.

2. Remove the used batteries and install 4 new AAA batteries following the diagram for 

correct polarity (see diagram).  We recommend high quality AAA alkaline batteries be used.

4. Replace the battery cover. Be sure to tighten the cover securely. This is necessary for meter to

be waterproof.

5. Dispose of the used batteries in accordance with your

local regulations.

3UHVV=(5221EXWWRQWRFRQ¿UPWKHPHWHUWXUQVRQ

    The meter is now ready for operation.

7. Meter will not work if battery orientation is incorrect.

To Install/Replace “AAA” Batteries:

TOP VIEW

Summary of Contents for Micro 7+

Page 1: ...stallation 34 Warranty 2 year 34 Reorder Information 35 USEPA Compliance 2 36 Kit Components 36 Visit us online at sensafe com micro7 for up to date product information NEW tests available IDEAL FOR D...

Page 2: ...OXGHG Battery Life WHVWV ZLWK DONDOLQH EDWWHULHV Electromagnetic Compliance PLWWHG QWHUIHUHQFH 1 EMC PPXQLW WR QWHUIHUHQFH 1 Waterproof Rating FHHGV 3 Weight QVWUXPHQW J R Dimensions QVWUXPHQW FP LQ e...

Page 3: ...DPD sulfate Anhydrous Na2HPO4 Anhydrous KH2PO4Na2 EDTA 1 5 33 4 64 0 1 1 1 5 33 4 64 0 1 1 Component Free Chlorine About The Accuracy Calibration Of The Micro 7 System OO WHVWV KDYH EHHQ FDOLEUDWHG XV...

Page 4: ...est Finally fill cell to capacity 4ml with the water sample RINSE AND FILL CELL WITH SAMPLE 4 Press and re press the MENU button until the display shows the parameter CL1 CL1 is also used for testing...

Page 5: ...asurement NOTE From testing in our lab water samples above 70 F 20 C generally reach a stabilized reading quicker then 2 minutes eXact Strip Micro CL DPD 1 DPD 3 DPD 4 Interferences part nos 486637 48...

Page 6: ...res the reading to be made after 2 minutes 2 minute wait is not necessary for Ozone measurements 6 Dip the eXact Strip Micro CL DPD 4 Part No 486670 into the CELL and immediately press READ This start...

Page 7: ...L Part No 486641 from the bottle before beginning the test Set the strip in a dry convenient place and recap the bottle immediately REMOVE STRIP 6 Dip the eXact Strip Micro AL Part No 486641 into the...

Page 8: ...Set the strip in a dry convenient place and recap the bottle immediately REMOVE STRIP 6 Dip the eXact Strip Micro CU Part No 486632 into the CELL and immediately press READ This starts the 20 SECOND...

Page 9: ...H EDFN DQG IRUWK PRWLRQ Remove and discard the after 1 on the display disappears 7KH FXUVRU ZLOO PRYH DFURVV WKH GLVSOD ZKLOH WKH PHWHU SUHSDUHV WR PHDVXUH WKH VDPSOH 5HFRUG UHVXOW GLVSOD HG WKLV UHVX...

Page 10: ...oss the display followed by 0 0 PPM The sample is ready for testing ZERO METER 2 Press the ZERO ON button to power the meter on the display will show all annunciators then the current MENU selection f...

Page 11: ...R7 result in the table below to determine the Cyanuric Acid concentration in ppm parts per million Example a TR7 result of 75 3 use only the 75 for the chart equals a Cyanuric Acid value of 6 8 ppm Re...

Page 12: ...splay shows the parameter TR7 SELECT TEST TR7 3 Rinse the CELL at least 3 times with the water sample you will be testing rinsing minimizes the potential for cross contamination from a previous test F...

Page 13: ...uals a Nitrate value of 3 40 ppm Record result After testing is completed rinse cell immediately to remove zinc dust USE TABLE 4 Press the ZERO ON button The cursor will move across the display follow...

Page 14: ...Use the small pipet provided to add 0 2ml of the sample to the water in the cell This is a 1 to 20 dilution of the pool water that is required to perform this test NOTE If you are using the more accur...

Page 15: ...LPSRUWDQW WR UHPRYH UHVLGXDO ZKLFK PD LQWHUIHUH Each eXact Strip Micro is valid for ONLY one test Discard strip after single use in regular refuse that is inaccessible to children and pets Each bottle...

Page 16: ...hat is using turbidity or precipitation chemistry for analysis Calcium Hardness and Cyanuric Acid eXact Photometer 2 Year Limited Warranty Registration of your eXact photometer must be received within...

Page 17: ...sion 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 486001 487200 486602 486641 486636 486629 486637 484051 486672 486638 486670 484054 486632...

Page 18: ...D 3 486638 25 eXact Strip Micro pH 486639 25 eXact Strip Micro Total Alkalinity 486641 25 eXact Strip Micro Calcium Hardness 486629 25 eXact Strip Micro Copper 486632 25 eXact Reagent Cyanuric Acid 48...

Page 19: ...Cell Cleaning Brush This Instruction Booklet Plastic Carrying Case eXact Micro 7 Kits y 1 0402x R 2 0 9989 0 0 0 5 1 0 1 5 2 0 2 5 3 0 3 5 4 0 4 5 5 0 5 5 6 0 0 0 0 5 1 0 1 5 2 0 2 5 3 0 3 5 4 0 4 5 5...

Page 20: ...6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 486001 487200 486602 486641 486637 486629 486637 484051 486672 486638 486670 484054 486632 486670 486637...

Page 21: ...u si le photom tre ne se remet pas z ro lorsque vous enfoncez le bouton ZERO ON il a besoin d tre nettoy Nettoyez comme suit remplissez la cellule d eau propre et frottez avec la brosse de haut en bas...

Page 22: ...in d emp cher tout r sidu KI de causer une interf rence il est tr s important de rincer le photom tre lorsque vous ex cutez une analyse de chlore libre DPD 1 apr s une analyse de chlore total DPD 3 ou...

Page 23: ...u ce que vous affichiez le param tre TR7 Choisissez l analyse TR7 3 Avec de l eau distill e ou d min ralis e rincez la cellule au moins trois fois afin de minimiser le potentiel de contamination crois...

Page 24: ...les 66 pour le graphique est gal une valeur de nitrate de 3 40 ppm Enregistrez le r sultat Apr s le test rincez la cellule imm diatement R f rez vous au tableau 4 Appuyez sur le bouton ZERO ON Le curs...

Page 25: ...ssez l analyse TR7 3 Rincez la cellule au moins 3 fois avec de l eau de l chantillon que vous allez tester le rin age minimise le risque potentiel de contamination crois e partir d un test pr c dent E...

Page 26: ...ar million Exemple un r sultat TR de 75 3 utilisez uniquement les 75 pour le graphique est gal une valeur d acide cyanurique de 6 8 ppm Enregistrez le r sultat Apr s le test rincez la cellule imm diat...

Page 27: ...str automatiquement dans le menu HR8 Apr s le test rincez la cellule imm diatement 5 Mettez le photom tre z ro Appuyez sur le bouton ZERO ON Le curseur se d placera jusqu afficher 0 00 PPM L chantillo...

Page 28: ...oins 3 fois avec de l eau de l chantillon que vous allez tester le rin age minimise le risque potentiel de contamination crois e partir d un test pr c dent Enfin remplissez la cellule capacit 4 ml ave...

Page 29: ...la bande eXact Strip Micro CU pi ce No 486632 dans le compartiment de la cellule et appuyez imm diatement sur le bouton READ Cette op ration d marre le compte rebours de 20 secondes Pendant ce temps...

Page 30: ...e num ro 486641 dans le compartiment de la cellule et appuyez imm diatement sur le bouton READ Cette op ration d marre le compte rebours de 20 secondes Pendant ce temps remuez doucement la bande dans...

Page 31: ...istr le Chlore Total avec la plus grandes des valeurs affich es par le photom tre REMARQUE La M thode Standard 4500 CI proc dure pour le Chlore Total requi re que la lecture du r sultat soit faite apr...

Page 32: ...n conformit vous devez compter 2 minutes et apr s faire la mesure REMARQUE D apr s nos tests en laboratoire les chantillons d eau au dessus de 20 C 70 F g n ralement atteignent un niveau stable de lec...

Page 33: ...ez tester le rin age minimise le risque potentiel de contamination crois e partir d un test pr c dent Enfin remplissez la cellule capacit 4 ml avec l chantillon d eau Trempez la bande eXact Strip Micr...

Page 34: ...enfonc e le bouton MENU pour faire d filer les r sultats enregistr s pour ce test en commen ant par le r sultat le plus r cent Le compteur affichera les 20 derni res lectures en commen ant par la plus...

Page 35: ...22 F Alimentation 4 piles alcalines AAA Non Incluse Dur e de vie des piles 2000 analyses avec piles alcalines Conformit mission parasitaire EN 61326 lectromagn tique CEM Immunit aux interf rences EN 6...

Page 36: ...INSI QUE POUR LES ANALYSES ENVIRONNEMENTALES ET DUCATIVES R vision 11 22 11 486691 K Kit Standard 486691 KP Kit Piscine et Spa 486691 WD Kit Forage Le Syst me photom trique de pointe eXact Micro 7 a t...

Reviews: