General description
3
The following values are monitored
-
Conductivity under the minimum limit value (only if the service valve is
opened)
-
The maximum conductivity limit value (from version 1.01.02 only if the
service valve is opened).
-
Inlet tank low level
-
Inlet tank high level
Step "No Service”
This step only appears in single filter systems.
The system is switched off when the filter cylinder is saturated.
Once the filter cylinder is exchanged, it can be switched on by pressing the external
key (press 1x).
Step "No Service 1+2”
If both filter cylinders shortly after one another produce water with too high
conductivity, the system is switched off. Once the filter cylinders are exchanged,
these must be switched on by pressing the external key (press 2x). Filter 1 will in
this case be switched on first.
During this step, conductivity is not monitored and the service valves remain closed.
Step "No Service 2+1”
If both filter cylinders shortly after one another produce water with too high
conductivity, the system is switched off. Once the filter cylinders are exchanged,
these can be switched on by pressing the external key (press 2x). Filter 2 will in
this case be switched on first.
During this step, conductivity is not monitored and the service valves remain closed.