DEPOLOX® Pool E flow cell
4. Installation
29
EN
10 Refit the housing cover.
Fig. 10
pH and ORP (Redox) single-rod
measuring chain with stand
A Sensor
B
Sealing cap
C
O-Ring
D Tank
E
Stand
4.18 Renewed start up
1 Prepare the sensors. See relevant instruc-
tions for the sensors.
2 Refit the sensors and connect the sensors
with DFMe electronics module (chapter
4.13 and 4.14).
3 Filling electrode-cleaning sand (chapter
4 For recommissioning, see instruction
manual “700 P electronics module".
4.19 Installing retrofit kits/extensions
4.19.1 Retrofitting sensor measuring
modules, total chlorine embrane
sensor TC3 CAN and module SiDi-
Sens conductivity measurement
1 Disconnect the electronics module from
the power supply. See instruction manual
“700 P Electronics module“.
2 Remove the housing cover of the flow
cell.
3 Take out all sensors from the mounting
hole in the cell body cover and unplug the
cable gland from DFMe electronics
module. Hold the cable while doing this as
it must not be allowed to twist.
4 Keep the sensors in a safe place (see inst-
ruction manual for sensors).
5 Disconnect CAN connection cable bet-
ween DFMe electronics module and 700 P
electronics module.
6 Remove DFMe electronics module from
the DEPOLOX® Pool E flow cell housing. To
do this, release the fixing screw on the
underside of DFMe electronics module.
7 Unhook DFMe electronics module from
the mounting hook (design chapter 4.13 -
Fig. 8) of DEPOLOX® Pool E flow cell. To do
this, raise DFMe electronics module and
carefully unhook it from the mounting
hook.
8 Open the housing of DFMe electronics
module. To do this, release the four cover
screws of DFMe electronics module.
9 Insert the optional sensor measuring
module cards in the provided slots (design
chapter 6.4).
10 When installing the SiDiSens LF module
and/or the membrane sensor, switch off
the CAN-balancing resistors (on the DFMe
electronics module) or set DIP switch
(Fig. 11) to the "OFF" position. Further-
more, the CAN extension socket (design
chapter 4.13 - Fig. 8 - item K) must also be
retrofitted. To do this, break out the
indent on the housing underside of DFMe
electronics module and fit the CAN exten-
sion socket. Check that all gasket inserts
are correctly mounted.
Fig. 11
Cutout of circuit board
DANGER
Risk of injury or death!
External voltages may still be connected
even if the operating voltage is switched off.
A
B
C
D
E
DIP
switch